Res Vet Sci. 2018 Jun;118:491-497. doi: 10.1016/j.rvsc.2018.05.002. Epub 2018 May 7.
Prevention of metabolic diseases in small ruminants may improve production efficiency and profitability, yet ewes carrying multiples or who are in poor body condition are at increased susceptibility to develop ketosis. This study evaluated the hand-held Nova Vet Meter to accurately detect β-hydroxybutyric acid (BHBA) concentrations in ewes and determined the percentage of ewes at moderate (0.8 to 1.5 mmol/L BHBA) and greatest (≥1.6 mmol/L BHBA) risk to develop ketosis during late gestation. To validate the Nova Vet Meter, BHBA concentrations of 104 paired blood samples were measured using the Nova Vet Meter and gold-standard laboratory analysis. Receiver operating characteristics were calculated. The accuracy and sensitivity of detecting BHBA concentrations at 0.8 to 1.5 mmol/L were 94.2% and 97.3%, respectively. The accuracy and sensitivity of detecting BHBA concentrations ≥ 1.6 mmol/L were 98.0% and 50.0%, respectively. Ewe body weight (BW), body condition score (BCS), and BHBA of 117 ewes from three flocks were determined weekly during the four weeks before parturition. During the last three weeks of gestation >20% of ewes were identified with moderate risk to develop ketosis. During the last four weeks of gestation, ewes carrying triplets had reduced BCS (P = 0.0002) and increased BHBA concentrations (P < 0.0001) compared with singleton and twin pregnancies. Ewe BHBA did not correlate with lamb birth weight (R = 0.003; P = 0.41). In conclusion, the Nova Vet Meter is suitable for sheep-side BHBA monitoring between 0.8 and 1.5 mmol/L, but further testing is necessary to evaluate BHBA readings ≥1.6 mmol/L.
预防小反刍动物的代谢疾病可以提高生产效率和盈利能力,但怀多胎或体况不佳的母羊更容易发生酮病。本研究评估了 Nova Vet Meter 手持设备来准确检测母羊血液中β-羟丁酸(BHBA)的浓度,并确定了在妊娠后期处于中度(0.8 至 1.5mmol/L BHBA)和最大(≥1.6mmol/L BHBA)风险的母羊比例,以发生酮病。为了验证 Nova Vet Meter,使用 Nova Vet Meter 和黄金标准实验室分析对 104 对血液样本的 BHBA 浓度进行了测量。计算了接收器工作特征。检测 0.8 至 1.5mmol/L 之间的 BHBA 浓度的准确性和灵敏度分别为 94.2%和 97.3%。检测≥1.6mmol/L 的 BHBA 浓度的准确性和灵敏度分别为 98.0%和 50.0%。在分娩前的四周内,每周对来自三个羊群的 117 只母羊进行了体重(BW)、体况评分(BCS)和 BHBA 的测定。在妊娠的最后三周,超过 20%的母羊有中度发展酮病的风险。在妊娠的最后四周,怀三胞胎的母羊的 BCS(P=0.0002)和 BHBA 浓度(P<0.0001)均低于单胎和双胎妊娠。母羊 BHBA 与羔羊初生重之间没有相关性(R=0.003;P=0.41)。总之,Nova Vet Meter 适合于 0.8 至 1.5mmol/L 之间的绵羊 BHBA 监测,但需要进一步测试来评估≥1.6mmol/L 的 BHBA 读数。