To evaluate the clinical and tomographic outcomes after implantation of a new intrastromal corneal ring segment (ICRS) with 140-degrees of arc in eyes with corneal ectasia.
We evaluated patients with corneal ectasia implanted with Ferrara 140° ICRS from April 2010 to February 2015. Outcome measures included preoperative and postoperative corrected distance visual acuity (CDVA), keratometry simulated (K) reading, tomographic astigmatism and asphericity. All patients were evaluated using the Pentacam Scheimpflug system.
The study evaluated 58 eyes. The mean follow-up was 16.81±10.8mo. The CDVA (logMAR) improved from 0.5±0.20 (20/60) to 0.3±0.21 (20/40) (<0.01). The average K reduced from 49.87±7.01 to 47.34±4.90 D (<0.01). The asphericity changed from -0.60±0.86 to -0.23±0.67 D (<0.01). The mean preoperative tomographic astigmatism decreased from -8.0±3.45 to -4.53±2.52 D (<0.01).
The new ICRS model with 140-degrees of arc effectively improve the visual acuity and reduce the high astigmatism usually found in patients with corneal ectasia.