Ghammraoui Bahaa, Badal Andreu, Glick Stephen J
Office of Science and Engineering Laboratories, CDRH, U.S. Food and Drug Administration, Silver Spring, MD, 20993-0002, USA.
Med Phys. 2018 Jun 3. doi: 10.1002/mp.13031.
Mammographic density of glandular breast tissue has a masking effect that can reduce lesion detection accuracy and is also a strong risk factor for breast cancer. Therefore, accurate quantitative estimation of breast density is clinically important. In this study, we investigate experimentally the feasibility of quantifying volumetric breast density with spectral mammography using a CdTe-based photon-counting detector.
To demonstrate proof-of-principle, this study was carried out using the single pixel Amptek XR-100T-CdTe detector. The total number of x rays recorded by the detector from a single pencil-beam projection through 50%/50% of adipose/glandular mass fraction-equivalent phantoms was measured. Material decomposition assuming two, four, and eight energy bins was then applied to characterize the inspected phantom into adipose and glandular using log-likelihood estimation, taking into account the polychromatic source, the detector response function, and the energy-dependent attenuation.
Measurement tests were carried out for different doses, kVp settings, and different breast sizes. For dose of 1 mGy and above, the percent relative root mean square (RMS) errors of the estimated breast density was measured below 7% for all three phantom studies. It was also observed that some decrease in RMS errors was achieved using eight energy bins. For 3 and 4 cm thick phantoms, performance at 40 and 45 kVp showed similar performance. However, it was observed that 45 kVp showed better performance for a phantom thickness of 6 cm at low dose levels due to increased statistical variation at lower photon count levels with 40 kVp.
The results of the current study suggest that photon-counting spectral mammography systems using CdTe detectors have the potential to be used for accurate quantification of volumetric breast density on a pixel-to-pixel basis, with an RMS error of less than 7%.
乳腺腺体组织的乳腺X线密度具有掩盖效应,会降低病变检测的准确性,同时也是乳腺癌的一个重要危险因素。因此,准确地定量估计乳腺密度在临床上具有重要意义。在本研究中,我们通过实验研究了使用基于碲化镉(CdTe)的光子计数探测器进行乳腺密度定量光谱乳腺成像的可行性。
为了证明原理,本研究使用单像素Amptek XR-100T-CdTe探测器进行。测量探测器从单个铅笔束投影穿过脂肪/腺体质量分数等效为50%/50%的体模所记录的X射线总数。然后应用假设为2、4和8个能量区间的物质分解,通过对数似然估计将检查的体模表征为脂肪和腺体,同时考虑多色源、探测器响应函数和能量依赖衰减。
针对不同剂量、千伏峰值(kVp)设置和不同乳腺尺寸进行了测量测试。对于1毫戈瑞及以上的剂量,在所有三项体模研究中,估计乳腺密度的相对均方根(RMS)误差百分比测量值均低于7%。还观察到使用8个能量区间时RMS误差有所降低。对于3厘米和4厘米厚的体模,40 kVp和45 kVp时的性能表现相似。然而,观察到在低剂量水平下,对于6厘米厚的体模,45 kVp表现出更好的性能,因为40 kVp在较低光子计数水平下统计变化增加。
当前研究结果表明,使用CdTe探测器的光子计数光谱乳腺成像系统有潜力用于在逐像素基础上准确量化乳腺体积密度,RMS误差小于7%。