In the present study, we compared the cyclic fatigue resistance (CFR) of Reciproc Blue, WaveOne Gold, and SmartTrack files in curved artificial canals.
Ninety new Reciproc Blue R25, WaveOne Gold Primary, and SmartTrack X1 files were tested in artificial canals with 45° and 60° angles of curvature. CFR was determined by recording the time to fracture in the artificial canals. Two-way analysis of variance was used to analyze the data.
In the canal with a 45° angle of curvature, no significant differences were observed amongst Reciproc Blue, WaveOne Gold, and SmartTrack (P > .05). In the canal with a 60° angle of curvature, Reciproc Blue and SmartTrack had a greater CFR than WaveOne Gold (P < .05); there was no significant difference between the Reciproc Blue and SmartTrack files (P > .05).
The results of the present study showed that Reciproc Blue and SmartTrack files exhibited greater CFR than WaveOne Gold only in canals with a 60° angle of curvature.