Scand J Trauma Resusc Emerg Med. 2018 Jun 5;26(1):47. doi: 10.1186/s13049-018-0510-5.
BACKGROUND/AIM: Early identification of lactate levels may have a large impact on triage classification and assist in identifying critically ill patients. A handheld device provides a rapid and timesaving measurement of lactate levels adapted to work in a prehospital care setting. I.e., the device is small, fast, and easy-to-use. The aim of this study was to evaluate the Accutrend Plus handheld lactate analyzer in comparison to the reference in-hospital method.
Patients triaged as minimum yellow according to the RETTS System (Rapid Emergency Triage and Treatment System) and transported to hospital by ambulance were selected and a written consent to participate was obtained prior to inclusion in the study. Capillary (CAP) and venous (VEN) blood were analyzed with Accutrend Plus (AP). Venous blood samples were analyzed at the local hospital laboratory (GEM premier 4000) within 20 min from sampling. All sampling was conducted by two registered nurses specially trained in prehospital care.
480 lactate measurements were performed in 160 patients. The mean difference between measurements in capillary blood compared with the reference method was 0.7 mmol/L and for venous blood 0.9 mmol/L. The limits of agreement from the Bland-Altman plot was - 0.9 to + 2.5 mmol/L and and - 0.1 to + 1.9 mmol/L, for CAP and VEN compared with GEM.
Our results shows low accuracy and low precision with VEN / CAP measurements of lactate compared to reference GEM.
背景/目的:早期识别乳酸水平可能对分诊分类产生重大影响,并有助于识别危重症患者。手持式设备可快速、省时地测量乳酸水平,适用于院前护理环境。即,该设备体积小、速度快、易于使用。本研究旨在评估 Accutrend Plus 手持式乳酸分析仪与参考院内方法相比的性能。
根据 RETTS 系统(快速紧急分诊和治疗系统)分诊为最低黄色级别的患者,并由救护车送往医院,在纳入研究之前获得书面同意书。使用 Accutrend Plus(AP)分析毛细血管(CAP)和静脉(VEN)血。静脉血样在采样后 20 分钟内由当地医院实验室(GEM premier 4000)进行分析。所有采样均由两名专门接受过院前护理培训的注册护士进行。
对 160 名患者进行了 480 次乳酸测量。与参考方法相比,毛细血管血测量的平均差值为 0.7mmol/L,静脉血为 0.9mmol/L。Bland-Altman 图的一致性界限为 CAP 和 VEN 与 GEM 相比分别为 -0.9 至 +2.5mmol/L 和 -0.1 至 +1.9mmol/L。
与参考 GEM 相比,我们的结果显示静脉血/毛细血管血乳酸测量的准确性和精密度均较低。