使用高千伏和附加滤过降低剂量进行腹部数字 X 射线摄影的质量评估。

Image quality assessment with dose reduction using high kVp and additional filtration for abdominal digital radiography.

机构信息

Department of Radiology and Research Institute of Radiology, Asan Medical Center, University of Ulsan, College of Medicine, Seoul, South Korea; Department of Medical Physics, Korea University, South Korea.

Department of Medical Physics, Korea University, South Korea.

出版信息

Phys Med. 2018 Jun;50:46-51. doi: 10.1016/j.ejmp.2018.05.007. Epub 2018 May 26.

Abstract

PURPOSE

Dose reduction using additional filters with high kilovoltage peak (kVp) for abdominal digital radiography has received much attention recently. We evaluated image quality with dose reduction in abdominal digital radiography by using high kVp and additional copper filters at a tertiary hospital.

METHODS

Between June 2016 and July 2016, 82 patients underwent abdominal digital radiography using 80 kVp in X-ray room 1 and 82 were imaged using 92 kVp with 0.1-mm copper filtration in X-ray room 2. The effective dose was calculated using a PC-based Monte Carlo program. Image quality of the abdominal radiography acquired in the two rooms was evaluated using a five-point ordinal scale, as well as the signal-to-noise and contrast-to-noise ratios.

RESULTS

The mean effective dose decreased by 25.8% and 25.7% for the supine and standing positions, respectively, when abdominal digital radiography using 92 kVp with 0.1-mm copper filtration was performed. In the 20 patients who performed abdominal digital radiography twice in each room, visual grading scores for visualisation of psoas outlines and kidney outlines are higher in room 1. However, there was no statistical significant difference of visual grading scores among the 124 patients who underwent only one abdominal radiography in the room 1 or 2 (P > 0.05).

CONCLUSIONS

Dose reduction for abdominal digital radiography can be achieved with comparable image quality by performing abdominal digital radiography using 92 kVp with 0.1-mm copper filtration, despite the higher AEC dose.

摘要

目的

最近,使用高千伏峰值(kVp)附加滤光片减少腹部数字放射摄影的剂量已受到广泛关注。我们在一家三级医院评估了使用高 kVp 和附加铜滤器进行腹部数字放射摄影的剂量减少的图像质量。

方法

2016 年 6 月至 7 月,在 X 射线室 1 中使用 80kVp 对 82 例患者进行了腹部数字放射摄影,在 X 射线室 2 中使用 92kVp 和 0.1mm 铜滤器对 82 例患者进行了成像。使用基于 PC 的蒙特卡罗程序计算有效剂量。使用五分制有序量表以及信噪比和对比噪声比评估两个房间获得的腹部射线照相的图像质量。

结果

当使用 92kVp 和 0.1mm 铜滤器进行腹部数字放射摄影时,仰卧位和站立位的平均有效剂量分别降低了 25.8%和 25.7%。在每个房间进行两次腹部数字放射摄影的 20 例患者中,房间 1 中腰大肌轮廓和肾脏轮廓的可视化视觉分级评分更高。然而,在仅在房间 1 或 2 进行一次腹部放射摄影的 124 例患者中,视觉分级评分之间没有统计学上的显著差异(P>0.05)。

结论

尽管自动曝光控制剂量较高,但使用 92kVp 和 0.1mm 铜滤器进行腹部数字放射摄影可以实现剂量减少,同时保持可比的图像质量。