在阿联酋迪拜卫生局,对进行数字胸部 X 射线成像的成年人进行额外铜过滤对辐射剂量和图像质量的影响。

Potentials of additional copper filtration on radiation dose and image quality for adults underwent digital chest X-ray imaging in Dubai Health Authority - UAE.

机构信息

PHC-Diagnostic Imaging Department, DAHC, Dubai, United Arab Emirates.

HQ-Diagnostic Imaging Department, DAHC, Dubai, United Arab Emirates; College of Medicine, Mohammed Bin Rashid University, Dubai, United Arab Emirates.

出版信息

Radiography (Lond). 2023 May;29(3):552-556. doi: 10.1016/j.radi.2023.03.005. Epub 2023 Mar 21.

Abstract

OBJECTIVE

To explore the potentials of adding copper (Cu) filter on image quality and patient dose of adult patients underwent chest X-ray examination METHODS: Patients were divided into four groups. Group 1, patients were exposed with no added Cu filter (standard or control), group 2 a 0.1 mm Cu filter was added, group 3 acquired with 0.2 mm Cu filter and group 4 performed with 0.3 mm Cu filter. Exposure index (EI), entrance surface dose (ESD) and dose area product (DAP) were recorded from the modality and retrospectively analyzed. The visual grading analysis score (VGAS) was used to evaluate image quality. Mann-Whitney T-Test and one-way ordinary ANOVA Test were used to evaluate statistical differences including gender-based findings.

RESULTS

EI, ESD and DAP data for a total of 784 patients (422 male and 362 female) that underwent indirect digital chest radiography exam were collected. Image quality was maintained when adding 0.1 mm Cu filter achieved with ∼19% DAP reduction. Female showed a significant DAP reduction comparing to male registered in the same group.

CONCLUSIONS

Reducing dose when using indirect digital chest radiography is possible with no trade-off on image quality. No loss of image quality was reported, images were broadly comparable.

IMPLICATIONS FOR PRACTICE

This study highlights the importance of utilizing the additional copper filter in digital chest radiography.

摘要

目的

探讨在成人胸部 X 射线检查中添加铜(Cu)滤片对图像质量和患者剂量的潜在影响。

方法

将患者分为四组。组 1 患者未添加 Cu 滤片(标准或对照),组 2 添加 0.1mm Cu 滤片,组 3 添加 0.2mm Cu 滤片,组 4 添加 0.3mm Cu 滤片。从设备中记录曝光指数(EI)、体表剂量(ESD)和剂量面积乘积(DAP),并进行回顾性分析。使用视觉分级分析评分(VGAS)评估图像质量。采用曼-惠特尼 T 检验和单因素方差分析检验评估包括基于性别的发现在内的统计学差异。

结果

共收集了 784 名(422 名男性和 362 名女性)接受间接数字化胸部 X 射线检查的患者的 EI、ESD 和 DAP 数据。当添加 0.1mm Cu 滤片时,图像质量得以维持,同时 DAP 降低了约 19%。在同一组中,女性的 DAP 明显低于男性。

结论

在不影响图像质量的情况下,使用间接数字化胸部 X 射线摄影可以降低剂量。没有报告图像质量下降的情况,图像广泛可比。

意义

本研究强调了在数字化胸部 X 射线摄影中使用附加铜滤片的重要性。