The Accuracy of Cone-Beam Computed Tomography for Evaluating Bone Density and Cortical Bone Thickness at the Implant Site: Micro-Computed Tomography and Histologic Analysis.

作者信息

Suttapreyasri Srisurang, Suapear Pradipat, Leepong Narit

机构信息

Department of Oral and Maxillofacial Surgery, Faculty of Dentistry, Prince of Songkla University, Songkhla, Thailand.

出版信息

J Craniofac Surg. 2018 Nov;29(8):2026-2031. doi: 10.1097/SCS.0000000000004672.

Abstract

The aim of this study was to evaluate the accuracy of cone-beam computed tomography (CBCT) for determining cortical thickness and the gray value, investigating its correlation with micro-computed tomography (CT) and histology analysis. Sixty-two bone samples from 4 anatomic regions of the jaw were analyzed. A radiographic surgical stent was used during CBCT and bone sample harvesting. The cortical thickness and gray value of the planned implant were evaluated by CBCT. Bone volumetric fractions, bone mineral density, and % porosity assessed by micro-CT and mineralized material by histology analysis from harvested bone samples were analyzed and assessed for the association with the CBCT using Pearson correlation. A correlation between cortical thickness measured from the CBCT and Micro-CT (r = 0.933, P < 0.01) was identified. There was no difference between gray values measured from the CBCT among regions, while bone density parameters from micro-CT and histologic analysis showed significant difference (P < 0.01) among regions. Bone density parameters from micro-CT and histologic analysis showed correlation with cortical thickness but not with the gray value. In conclusion, CBCT is highly accurate in linear measurements and demonstrated correlation with genuine bone density. However, the gray value could not demonstrate the true bone density according to a low correlation to bone density variable measured from micro-CT and histologic analysis.

摘要

本研究的目的是评估锥形束计算机断层扫描(CBCT)在确定皮质骨厚度和灰度值方面的准确性,研究其与微型计算机断层扫描(CT)及组织学分析的相关性。对取自颌骨4个解剖区域的62块骨样本进行了分析。在CBCT扫描及采集骨样本过程中使用了放射外科支架。通过CBCT评估计划植入部位的皮质骨厚度和灰度值。对采集的骨样本通过微型CT评估骨体积分数、骨矿物质密度和孔隙率百分比,并通过组织学分析评估矿化物质,然后使用Pearson相关性分析评估其与CBCT的相关性。结果发现,CBCT测量的皮质骨厚度与微型CT测量结果之间存在相关性(r = 0.933,P < 0.01)。CBCT测量的各区域灰度值之间无差异,而微型CT和组织学分析的骨密度参数在各区域间存在显著差异(P < 0.01)。微型CT和组织学分析的骨密度参数与皮质骨厚度相关,但与灰度值无关。总之,CBCT在线性测量方面高度准确,且与实际骨密度存在相关性。然而,根据其与微型CT和组织学分析测量的骨密度变量的低相关性,灰度值无法反映真实的骨密度。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验