Ruiz-Martinez S, Volpe G, Vannuccini S, Cavallaro A, Impey L, Ioannou C
Obstetrics Department, Aragon Institute of Health Research, IIS Aragón, Hospital Clínico Universitario Lozano Blesa, Zaragoza, Spain.
Nuffield Department of Women's and Reproductive Health, University of Oxford, John Radcliffe Hospital, Oxford, UK.
J Matern Fetal Neonatal Med. 2020 Feb;33(3):421-426. doi: 10.1080/14767058.2018.1494711. Epub 2018 Sep 10.
To validate an objective scoring system for middle cerebral artery (MCA) pulsed wave Doppler images. From an image database of routine 36-week scans, a random sample of MCA Doppler images was selected. Two reviewers rated the images subjectively as acceptable or unacceptable. Subsequently they used an objective 6-point image scoring system and awarded one point for each of the following: (1) anatomical site, (2) magnification, (3) angle of insonation, (4) image clarity, (5) sweep speed adjustment, and (6) velocity scale and baseline adjustment. Image scores 4-6 were defined as good quality whereas 0-3 as poor. The subjective and objective agreement between the two reviewers was compared using the adjusted Kappa statistic. A total of 124 images were assessed. Using objective scoring the agreement rate between reviewers increased to 91.9% (κ = 0.839) compared to subjective agreement 75.8% (κ = 0.516). The agreement for each criterion was: anatomical site 91.1% (κ = 0.823), magnification 95.2% (κ = 0.903), clarity 83.9% (κ = 0.677), angle 96.0% (κ = 0.919), sweep speed 98.4% (κ = 0.968), and velocity scale and baseline 94.4% (κ = 0.887). Objective assessment of MCA Doppler images using a 6-point scoring system has greater interobserver agreement than subjective assessment and could be used for MCA Doppler quality assurance.
验证一种用于大脑中动脉(MCA)脉冲波多普勒图像的客观评分系统。从常规36周扫描的图像数据库中,随机抽取MCA多普勒图像样本。两名审阅者主观地将图像评定为可接受或不可接受。随后,他们使用一种客观的6分图像评分系统,对以下各项分别给予一分:(1)解剖部位,(2)放大倍数,(3)探测角度,(4)图像清晰度,(5)扫描速度调整,以及(6)速度标尺和基线调整。图像分数4 - 6被定义为高质量,而0 - 3为低质量。使用调整后的卡帕统计量比较两名审阅者之间的主观和客观一致性。共评估了124幅图像。与主观一致性75.8%(κ = 0.516)相比,使用客观评分时审阅者之间的一致率提高到了91.9%(κ = 0.839)。每个标准的一致性为:解剖部位91.1%(κ = 0.823),放大倍数95.2%(κ = 0.903),清晰度83.9%(κ = 0.677),角度96.0%(κ = 0.919),扫描速度98.4%(κ = 0.968),以及速度标尺和基线94.4%(κ = 0.887)。使用6分评分系统对MCA多普勒图像进行客观评估比主观评估具有更高的观察者间一致性,可用于MCA多普勒质量保证。