Suppr超能文献

Micra™无导线起搏器引起的亚临床心脏穿孔

Subclinical cardiac perforation caused by a Micra™ leadless pacemaker.

作者信息

Togashi Ikuko, Sato Toshiaki, Hoshida Kyoko, Soejima Kyoko

机构信息

Cardiology Kyorin University School of Medicine Mitaka Japan.

Division of Advanced Arrhythmia Management Kyorin University School of Medicine Mitaka Japan.

出版信息

J Arrhythm. 2018 Apr 6;34(3):326-328. doi: 10.1002/joa3.12052. eCollection 2018 Jun.

Abstract

A subclinical cardiac perforation by a device cup of the Micra™ transcatheter pacing system was suspected in a 78-year-old woman. During the procedure, the device cup was placed on the septum. The contrast media was injected before device deployment and remained outside of the myocardium. Later, a cardiac computed tomography scan visualized a protruded diverticular structure on the right ventricle. The contrast material remained in a pouch within the pericardium. To ensure the device is oriented away from the border between the right ventricular septum and the free wall, right anterior oblique view should be carefully reviewed before deployment.

摘要

一名78岁女性疑似发生了Micra™经导管起搏系统的设备杯导致的亚临床心脏穿孔。在手术过程中,设备杯放置在室间隔上。在设备植入前注射了造影剂,造影剂未进入心肌。后来,心脏计算机断层扫描显示右心室有一个突出的憩室结构。造影剂留在心包内的一个囊袋中。为确保设备远离右心室间隔和游离壁之间的边界,在植入前应仔细查看右前斜位视图。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/32d7/6009778/b684a95dd0ea/JOA3-34-326-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验