Pinto-Sinai G, Brewster J, Roberts H
Oper Dent. 2018 Sep/Oct;43(5):E266-E272. doi: 10.2341/17-381-L. Epub 2018 Jun 28.
The purpose of this evaluation was to evaluate the linear coefficient of thermal expansion (LCTE) of 12 conventional glass ionomer (GIC) and four resin-modified glass ionomer (RMGI) restorative materials.
GIC and RMGI specimens (2 mm × 5 mm × 5 mm) were fabricated (n=12) following manufacturer instructions and were placed in 0.2M phosphate-buffered saline and stored at 37°C and 98% humidity for one week. Specimens had LCTE determined with a thermomechanical analysis (TMA) unit using a 15°C-50°C heating cycle as well as a 50°C-15°C cooling cycle at a 5°C/min rate, using a 3-mm ball-point probe under 0.02 N probe pressure with all specimens kept saturated with PBS using a specially designed quartz container. Each specimen was tested three times, with the mean representing the specimen LCTE. Mean results between specimen heating and cooling were compared with paired Wilcoxon sign rank test, while results between materials were compared with Kruskal-Wallis/Dunn's ( α=0.05).
GIC LCTE ranged from approximately 5°C to 20°C ppm °K, while the RMGI LCTE ranged from approximately 25°C to 47°C ppm °K. With some exception, the LCTE during cooling displayed a greater trend.
Under moisture conditions similar to the oral cavity, GIC materials overall had LCTE values closer to that reported for tooth structure. RMGI materials displayed higher values, which was thought to be related to the amount of resin in the matrix. A generally greater LCTE trend with cooling for all materials was noted, but the small magnitude of the difference is presently thought to be of minor clinical significance.
本评估旨在测定12种传统玻璃离子水门汀(GIC)和4种树脂改性玻璃离子水门汀(RMGI)修复材料的热膨胀线性系数(LCTE)。
按照制造商说明制作GIC和RMGI样本(2毫米×5毫米×5毫米)(n = 12),将其置于0.2M磷酸盐缓冲盐水中,在37°C和98%湿度条件下储存一周。使用热机械分析仪(TMA)测定样本的LCTE,加热循环为15°C至50°C,冷却循环为50°C至15°C,速率为5°C/分钟,使用3毫米圆珠探针,探针压力为0.02 N,所有样本使用专门设计的石英容器保持在PBS中饱和状态。每个样本测试三次,平均值代表样本的LCTE。样本加热和冷却之间的平均结果采用配对Wilcoxon符号秩检验进行比较,材料之间的结果采用Kruskal-Wallis/Dunn检验(α = 0.05)。
GIC的LCTE范围约为5°C至20°C ppm °K,而RMGI的LCTE范围约为25°C至47°C ppm °K。除了一些例外情况,冷却过程中的LCTE呈现出更大的趋势。
在类似于口腔的湿度条件下,GIC材料的LCTE值总体上更接近牙齿结构的报告值。RMGI材料显示出更高的值,这被认为与基质中的树脂含量有关。注意到所有材料在冷却时LCTE总体上有更大的趋势,但目前认为差异幅度较小,临床意义不大。