Suppr超能文献

评估电子术中放疗(ELIOT)作为心脏植入式电子设备(CIED)患者部分乳房照射的可行性的剂量学研究。

Dosimetric study to assess the feasibility of intraoperative radiotherapy with electrons (ELIOT) as partial breast irradiation for patients with cardiac implantable electronic device (CIED).

机构信息

Medical Physics Unit, European Institute of Oncology, Milan, Italy.

Radiation Oncology Division, European Institute of Oncology, Milan, Italy.

出版信息

Breast Cancer Res Treat. 2018 Oct;171(3):693-699. doi: 10.1007/s10549-018-4878-8. Epub 2018 Jul 5.

Abstract

PURPOSE

To report in-vivo dosimetry in the infraclavicular region, a potential site of a cardiac implantable electronic device (CIED) and to evaluate the absorbed dose from intraoperative radiotherapy with electrons (ELIOT).

METHODS

27 non-cardiopathic breast cancer (BC) patients without CIED received quadrantectomy and ELIOT as partial breast irradiation. Before delivering ELIOT, two catheters, each containing eight thermoluminescent dosimeters (TLDs), were positioned in the infraclavicular region. TLDs internal catheter was located deep in the tumor bed while the external catheter was placed on patient's skin.

RESULTS

Data were available for 24/27 patients. The absorbed doses were referred to the dose of 21 Gy. Values measured by the external catheter were low, although statistically significant higher doses were found close to the applicator (mean values 0.26-0.49 Gy). External TLD doses in proximity of the applicator were lower than those detected by their internal counterparts. Values measured by the internal catheter TLDs varied according to the distance from the applicator while no correlation with tumor site and beam energy was found. The distance from the applicator to deliver < 2 Gy to a CIED was 2 cm, while from 2.5 cm the dose measured in all the patients became negligible.

CONCLUSIONS

This dosimetric study provided data to support the clinical use of ELIOT in BC patients having CIEDs as long as the suggested minimum safe distance of 2.5 cm is taken from the RT field in case of ELIOT single dose of 21 Gy, in the energy range of 6-10 MeV.

摘要

目的

报告锁骨下区域的体内剂量学情况,锁骨下区域是心脏植入式电子设备(CIED)的潜在部位,并评估电子术中放射治疗(ELIOT)的吸收剂量。

方法

27 例无 CIED 的非心脏病乳腺癌(BC)患者接受象限切除术和 ELIOT 作为部分乳房照射。在实施 ELIOT 之前,将两个导管(每个导管包含八个热释光剂量计(TLD))置于锁骨下区域。TLD 内部导管位于肿瘤床深处,而外部导管置于患者皮肤表面。

结果

24/27 例患者的数据可用。所测吸收剂量参考 21 Gy 的剂量。尽管靠近敷贴器处的剂量较高,但外部导管所测剂量较低(平均值为 0.26-0.49 Gy)。外部 TLD 剂量在敷贴器附近低于内部 TLD 所测剂量。内部导管 TLD 所测剂量随与敷贴器的距离而变化,与肿瘤部位和射束能量无关。要使 CIED 接收到 <2 Gy 的剂量,需距离敷贴器 2 cm,而从 2.5 cm 起,所有患者的测量剂量变得微不足道。

结论

该剂量学研究提供了数据支持,只要在 ELIOT 单次 21 Gy(6-10 MeV 能量范围内)的情况下,从 RT 场中留出建议的 2.5 cm 的最小安全距离,就可以在有 CIED 的 BC 患者中安全使用 ELIOT。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验