比较加权滤波反投影全剂量腹部 CT 扫描重建与改良正弦图确认迭代重建(SAFIRE*)半剂量扫描重建的图像噪声和图像质量。

Comparison of image noise and image quality between full-dose abdominal computed tomography scans reconstructed with weighted filtered back projection and half-dose scans reconstructed with improved sinogram-affirmed iterative reconstruction (SAFIRE*).

机构信息

Department of Radiology, Vancouver General Hospital, University of British Columbia, 3350-950 W 10th Avenue, Vancouver, BC, V5Z 1M9, Canada.

Siemens Medical Solutions, Malvern, PA, USA.

出版信息

Abdom Radiol (NY). 2019 Jan;44(1):355-361. doi: 10.1007/s00261-018-1687-9.

Abstract

PURPOSE

To retrospectively compare the image noise, signal-to-noise ratio (SNR), and subjective image quality between CT images acquired with a dual-source, split-dose imaging protocol reconstructed at full and half doses with weighted filtered back projection (wFBP) and an improved sinogram-affirmed iterative reconstruction algorithm (SAFIRE*).

METHODS

Fifty-three consecutive patients underwent contrast-enhanced CT of the abdomen using a standardized dual-source, single energy CT protocol. Half-dose images were retrospectively generated using data from one detector only. Full-dose datasets were reconstructed with wFBP, while half-dose datasets were reconstructed with wFBP and SAFIRE* strengths 1-5. Region of interest analysis was performed to assess SNR and noise. Diagnostic acceptability, subjective noise, and spatial resolution were graded on a 10-point scale by two readers. Statistical analysis was carried out with repeated measures analysis of variance, Wilcoxon signed rank test, and Cohen's κ test.

RESULTS

With the increasing strengths of SAFIRE*, a progressive reduction in noise and increase in SNR (p < 0.01) was observed. There was a statistically significant decrease in objective noise and increase in SNR in half-dose SAFIRE* strength 4 and 5 reconstructions compared to full-dose reconstructions using wFBP (p < 0.01). Qualitative analysis revealed a progressive increase in diagnostic acceptability, decrease in subjective noise and increase in spatial resolution for half-dose images reconstructed with the increasing strengths of SAFIRE* (p < 0.01).

CONCLUSIONS

Half-dose CT images reconstructed with SAFIRE* at strength 4 and 5 have superior image quality compared to full-dose images reconstructed with wFBP. SAFIRE* potentially allows dose reductions in the order of 50% over wFBP.

摘要

目的

回顾性比较双源、分剂量成像方案采集的 CT 图像的图像噪声、信噪比(SNR)和主观图像质量,这些图像分别使用全剂量和半剂量重建的加权滤波反投影(wFBP)和改良的正弦图确认迭代重建算法(SAFIRE*)进行重建。

方法

53 例连续患者使用标准化双源、单能量 CT 方案进行腹部增强 CT 检查。仅使用一个探测器生成半剂量图像。全剂量数据集使用 wFBP 进行重建,而半剂量数据集则使用 wFBP 和 SAFIRE*强度 1-5 进行重建。进行感兴趣区域分析以评估 SNR 和噪声。由两位读者使用 10 分制对诊断可接受性、主观噪声和空间分辨率进行评分。使用重复测量方差分析、Wilcoxon 符号秩检验和 Cohen's κ 检验进行统计分析。

结果

随着 SAFIRE强度的增加,噪声逐渐降低,SNR 逐渐升高(p<0.01)。与全剂量 wFBP 重建相比,半剂量 SAFIRE强度 4 和 5 重建的客观噪声降低,SNR 升高,差异具有统计学意义(p<0.01)。定性分析显示,随着 SAFIRE*强度的增加,半剂量图像的诊断可接受性逐渐提高,主观噪声逐渐降低,空间分辨率逐渐提高(p<0.01)。

结论

与全剂量 wFBP 重建相比,SAFIRE强度 4 和 5 重建的半剂量 CT 图像具有更高的图像质量。SAFIRE 可能使剂量减少约 50%。