比较 Spot 和 SW800 筛查仪在中国儿童中的性能。

Compared performance of Spot and SW800 photoscreeners on Chinese children.

机构信息

Department of Strabismus and Pediatric Ophthalmology, Tianjin Medical University Eye Hospital, Tianjin, China.

Department of Ophthalmology, Emory University School of Medicine, Atlanta, Georgia, USA.

出版信息

Br J Ophthalmol. 2019 Apr;103(4):517-522. doi: 10.1136/bjophthalmol-2018-311885. Epub 2018 Jul 9.

Abstract

PURPOSE

To evaluate the effectiveness of Spot photoscreener and SW800 vision screener in detecting amblyopia risk factors in Chinese children between 4 and 6 years of age.

METHODS

One hundred and thirteen children (226 eyes) underwent complete ophthalmologic examination, cycloplegic retinoscopy refraction, prism cover tests and photoscreen using both Spot (v2.1.4) and SW800 (v1.0.1.0) photoscreeners. The agreement of results obtained from photoscreener and retinoscopy was evaluated by paired t-test as well as Pearson correlation test. The sensitivity and specificity of detecting amblyopia risk factors were calculated based on the American Association of Pediatric Ophthalmology and Strabismus 2013 guidelines. The overall effectiveness of detecting amblyopia risk factors by using either photoscreener was analysed by receiver operating characteristic (ROC) curves.

RESULTS

A strong linear agreement was observed between Spot and retinoscopy (p<0.01) in aspects of spherical equivalent (SE, Pearson's r=0.95), dioptre sphere (DS, r=0.97), dioptre cylinder (DC, r=0.84) and horizontal deviation (Hdev, r=0.91), with overall -0.17 D myopic shift of SE. Significant correlation was also shown between SW800 and retinoscopy (p<0.01) in aspects of SE (r=0.90), DS (r=0.93), DC (r=0.82) and Hdev (r=0.80), with overall -0.12 D myopic shift of SE. The overall sensitivity and specificity in detecting amblyopia risk factors were 94.0% and 80.0% for Spot and 88.8% and 81.1% for SW800.

CONCLUSION

The measurements of Spot and SW800 photoscreener showed a strong agreement with cycloplegic retinoscopy refraction and prism cover tests. The performance of both screeners in detecting individual amblyopia risk factors is satisfactory. ROC analysis indicates that the Spot and SW800 performed very similarly in detecting amblyopia risk factors.

摘要

目的

评估 Spot 视力筛查仪和 SW800 视力筛查仪在检测 4 至 6 岁中国儿童弱视危险因素方面的有效性。

方法

对 113 名儿童(226 只眼)进行了全面眼科检查、睫状肌麻痹验光、棱镜遮盖试验和使用 Spot(v2.1.4)和 SW800(v1.0.1.0)视力筛查仪进行的照片筛查。通过配对 t 检验和 Pearson 相关检验评估照片筛查仪和视网膜镜检查结果的一致性。根据美国儿科学会眼科学和斜视学分会 2013 年的指南,计算出检测弱视危险因素的敏感性和特异性。通过受试者工作特征(ROC)曲线分析使用任何一种照片筛查仪检测弱视危险因素的整体效果。

结果

在球镜等效(SE,Pearson's r=0.95)、屈光度球镜(DS,r=0.97)、屈光度圆柱镜(DC,r=0.84)和水平偏差(Hdev,r=0.91)方面,Spot 和视网膜镜之间观察到很强的线性一致性,SE 的总体远视 0.17 D。SW800 和视网膜镜之间也显示出显著的相关性(p<0.01),SE(r=0.90)、DS(r=0.93)、DC(r=0.82)和 Hdev(r=0.80),SE 的总体远视 0.12 D。Spot 检测弱视危险因素的总体敏感性和特异性分别为 94.0%和 80.0%,SW800 分别为 88.8%和 81.1%。

结论

Spot 和 SW800 视力筛查仪的测量结果与睫状肌麻痹视网膜镜验光和棱镜遮盖试验具有很强的一致性。两种筛查仪在检测单个弱视危险因素方面的性能均令人满意。ROC 分析表明,Spot 和 SW800 在检测弱视危险因素方面表现非常相似。