IEEE Trans Ultrason Ferroelectr Freq Control. 2018 Jul;65(7):1102-1110. doi: 10.1109/TUFFC.2018.2827241.
Several challenges currently prevent the use of Doppler echocardiography to assess blood flow in the coronary arteries. Due to the anatomy of the coronary tree, out-of-plane flow and high beam-to-flow angles easily occur. Transit-time broadening in regions with high velocities leads to overestimation of the maximum velocity envelope, which is a standard clinical parameter for flow quantification. In this paper, a commercial ultrasound system was locally modified to perform trans-thoracic, 3-D high frame-rate imaging of the coronary arteries. The imaging sequence was then combined with 3-D tracking Doppler for retrospective estimation of maximum velocities. Results from simulations showed that 3-D tracking Doppler delivers sonograms with better velocity resolution and spectral SNR compared to conventional pulsed wave (PW) Doppler. Results were confirmed using in vitro recordings. Further simulations based on realistic coronary flow data showed that 3-D tracking Doppler can provide improved performance compared to PW Doppler, suggesting a potential benefit to patients. In vivo feasibility of the method was also shown in a healthy volunteer.
目前,有几个挑战阻碍了多普勒超声心动图在冠状动脉血流评估中的应用。由于冠状动脉树的解剖结构,很容易出现平面外血流和高束流角度。在高速度区域,渡越时间展宽会导致最大速度包络的高估,这是血流定量的一个标准临床参数。在本文中,对商业超声系统进行了局部修改,以进行经胸、3-D 高帧率冠状动脉成像。然后将成像序列与 3-D 跟踪多普勒相结合,用于最大速度的回顾性估计。模拟结果表明,与传统的脉冲波(PW)多普勒相比,3-D 跟踪多普勒提供的超声图像具有更好的速度分辨率和谱 SNR。通过离体记录进行了验证。进一步基于真实的冠状动脉血流数据的模拟表明,与 PW 多普勒相比,3-D 跟踪多普勒可以提供更好的性能,这表明对患者有潜在的益处。在健康志愿者中也证明了该方法的体内可行性。