IEEE Trans Biomed Eng. 2019 Jan;66(1):237-245. doi: 10.1109/TBME.2018.2830303. Epub 2018 Apr 27.
Reported studies pertaining to needle guidance suggest that tissue impedance available from neuromonitoring systems can be used to discriminate nerve tissue proximity. In this pilot study, the existence of a relationship between intraoperative electrical impedance and tissue density, estimated from computer tomography (CT) images, is evaluated in the mastoid bone of in vivo sheep. In five subjects, nine trajectories were drilled using an image-guided surgical robot. Per trajectory, five measurement points near the facial nerve were accessed and electrical impedance was measured (≤1 KHz) using a multipolar electrode probe. Micro-CT was used postoperatively to measure the distances from the drilled trajectories to the facial nerve. Tissue density was determined from coregistered preoperative CT images and, following sensitivity field modeling of the measuring tip, tissue resistivity was calculated. The relationship between impedance and density was determined for 29 trajectories passing or intersecting the facial nerve. A monotonic decrease in impedance magnitude was observed in all trajectories with a drill axis intersecting the facial nerve. Mean tissue densities intersecting with the facial nerve (971-1161 HU) were different (p <0.01) from those along safe trajectories passing the nerve (1194-1449 HU). However, mean resistivity values of trajectories intersecting the facial nerve (14-24 Ωm) were similar to those of safe passing trajectories (17-23 Ωm). The determined relationship between tissue density and electrical impedance during neuromonitoring of the facial nerve suggests that impedance spectroscopy may be used to increase the accuracy of tissue discrimination, and ultimately improve nerve safety distance assessment in the future.
已有研究表明,神经监测系统提供的组织阻抗可用于区分神经组织的接近程度。在这项初步研究中,评估了活体绵羊乳突骨中术中电阻抗与从计算机断层扫描 (CT) 图像估计的组织密度之间的关系。在 5 个对象中,使用图像引导手术机器人钻了 9 个轨迹。对于每个轨迹,在面神经附近访问了 5 个测量点,并使用多极电极探头测量了电阻抗(≤1 kHz)。术后使用微 CT 测量从钻轨迹到面神经的距离。从术前 CT 图像配准确定组织密度,并在测量尖端进行敏感场建模后,计算组织电阻率。确定了 29 条穿过或与面神经交叉的轨迹的阻抗与密度之间的关系。在所有与面神经交叉的钻轴轨迹中,观察到阻抗幅度呈单调下降。与神经安全通过轨迹(1194-1449 HU)相比,与面神经相交的组织密度平均值(971-1161 HU)不同(p <0.01)。然而,与面神经相交的轨迹的平均电阻率值(14-24 Ωm)与安全通过轨迹(17-23 Ωm)相似。在面神经神经监测过程中确定的组织密度与电阻抗之间的关系表明,阻抗光谱分析可能用于提高组织区分的准确性,并最终提高未来神经安全距离评估的准确性。