Abstract

BACKGROUND

Hallux valgus disease is a common deformity of the forefoot. There are currently more than 100 surgical approaches for operative treatment. Because hypermobility of the first tarsometatarsal joint is considered to be causal for hallux valgus disease, fusion of the tarsometatarsal joint is an upcoming surgical procedure. Despite the development of new and increasingly stable fixation devices like different locking plates, malunion rates have been reported in 5 to 15% of cases.

METHODS

Biomechanical comparison of three commonly used fixation devices (a dorsal locking plate, a plantar locking plate, and an intramedullary fixation device) was performed by weight-bearing simulation tests on synthetic bones. Initial compression force and stiffness during simulation of postoperative weight-bearing were analysed.

RESULTS

Fixation of the first tarsometatarsal joint with the plantar plate combination demonstrated a higher stiffness compared to fixation with the intramedullary implant or the medial locking plate. The intramedullary device provided the highest initial compression force. Failure was detected in the following ranking: (1) the angle-stable intramedullary fixation device, (2) the medial located plate, and (3) the plantar locking plate.

CONCLUSION

The intramedullary device demonstrated the highest initial compression force of the three tested implants. The plantar locking plate showed the best overall stability during weight-bearing simulation. Further clinical research is necessary to analyse if the intramedullary fixation device needs a longer period of non-weight-bearing to reach a better non-union rate compared to the plantar locking plate.

摘要

背景

足拇外翻是一种常见的前足畸形。目前有超过 100 种手术方法可用于治疗。由于第一跖跗关节的过度活动被认为是导致足拇外翻的原因,因此跖跗关节融合术是一种即将出现的手术方法。尽管新的和越来越稳定的固定装置(如不同的锁定板)的发展,但仍有 5%至 15%的病例报告出现愈合不良。

方法

通过对合成骨进行负重模拟试验,对三种常用固定装置(背侧锁定板、跖侧锁定板和髓内固定装置)进行了生物力学比较。分析了模拟术后负重时的初始压缩力和刚度。

结果

与髓内植入物或内侧锁定板固定相比,跖侧钢板组合固定第一跖跗关节具有更高的刚度。髓内装置提供了最高的初始压缩力。按以下顺序出现失效:(1)角稳定髓内固定装置,(2)内侧定位板,(3)跖侧锁定板。

结论

在三种测试的植入物中,髓内装置表现出最高的初始压缩力。在负重模拟过程中,跖侧锁定板表现出最佳的整体稳定性。需要进一步的临床研究来分析与跖侧锁定板相比,髓内固定装置是否需要更长的非负重期才能达到更好的愈合率。