Burstal Richard J, Threlfo Steven J
Department of Anaesthesia, John Hunter Hospital, Newcastle, New South Wales, Australia.
Paediatr Anaesth. 2018 Sep;28(9):774-779. doi: 10.1111/pan.13447. Epub 2018 Jul 13.
Most anesthetic ventilators are designed to cope with a wide range of patient sizes and may lack precision at the lowest end of the minute volume scale. Neonatal intensive care ventilators on the other hand are designed specifically for this patient group, but are not able to deliver volatile anesthesia.
We aimed to adapt the neonatal ventilator currently in use in our institution to deliver sevoflurane by incorporating a vaporizer and a scavenging system.
We used a Diamedica draw-over vaporizer incorporated into the ventilator circuit and a custom designed open interface scavenging system. A number of safety measures are described to ensure that this equipment is correctly inserted into the circuit.
Bench testing revealed that the vaporizer output is linear and stable within the circuit flow range 4-8 L/min in all modes except high frequency oscillation where concentrations are not predictable. The scavenging system was found to be effective and did not affect volumes, pressures or waveforms when ventilating a test lung over a wide range of flows and pressures. This remained the case over the full range of scavenger flow adjustment.
The addition of a Diamedica vaporizer to a Fabian neonatal ventilator was shown in bench testing to provide stable, linear vapor concentrations without compromise of ventilator function. The system should not be used in high frequency oscillation mode because concentrations will exceed those expected and will not maintain a linear relationship with the vaporizer setting.
大多数麻醉呼吸机旨在应对各种体型的患者,在分钟通气量范围的最低端可能缺乏精确性。另一方面,新生儿重症监护呼吸机是专门为这一患者群体设计的,但无法提供挥发性麻醉。
我们旨在通过整合蒸发器和清除系统,对我院目前使用的新生儿呼吸机进行改造,以输送七氟烷。
我们在呼吸机回路中使用了一个Diamedica旁流式蒸发器和一个定制设计的开放式接口清除系统。描述了一些安全措施,以确保该设备正确插入回路。
台架测试表明,除高频振荡模式下浓度不可预测外,在所有模式下,蒸发器输出在4-8升/分钟的回路流量范围内呈线性且稳定。在对测试肺进行大范围的流量和压力通气时,清除系统被证明是有效的,并且不会影响容量、压力或波形。在清除器流量调整的全范围内均如此。
台架测试表明,在Fabian新生儿呼吸机上添加Diamedica蒸发器可提供稳定的线性蒸汽浓度,且不影响呼吸机功能。该系统不应在高频振荡模式下使用,因为浓度将超过预期,并且与蒸发器设置不会保持线性关系。