Wu Zhi-Qi, Xu Hua-Guo
Department of Laboratory Medicine, the First Affiliated Hospital of Nanjing Medical University, Nanjing, China.
J Clin Lab Anal. 2019 Jan;33(1):e22618. doi: 10.1002/jcla.22618. Epub 2018 Jul 13.
The purpose of our study was to investigate whether the storage time and temperature of internal quality control (IQC) material influence the result of ACTH in IQC measurements.
Five levels of IQC materials from two manufacturers were tested through the precision of ACTH, the three freeze/thaw cycles, and the storage time and temperature to evaluate the stability of IQC material. All commercial control materials were simultaneously tested three times a day for five consecutive days.
Total precision of three levels of Bio-Rad IQC sera was 13.93%, 16.45%, and 15.98%, respectively, but repeatability was <2%. The concentration of ACTH decreased by 30%-50% after 3 freeze/thaw cycles. At room temperature, the concentration of ACTH from 3 levels decreased by 16.60%, 17.98%, and 17.20%, respectively, after 0.5 hours, and 70.54%, 74.36%, and 72.03%, respectively, after 4 hours. However, after 2 hours of storage at 4°C, the decline in the measured ACTH IQC was 8.04%, 11.84%, and 10.11%, respectively. Total precision of Roche IQC was 1.17% and 1.08%, respectively. After 3 freeze/thaw cycles, the concentration of ACTH decreased <5%. After 4 hours, the change of ACTH still steadied within 5% both at the room temperature and at 4°C.
Roche is a better choice for ACTH of IQC material in Elecsys immunoassay system in our study. If Bio-Rad control materials be used in Elecsys immunoassay system for ACTH IQC testing material, it should be stored at 4°C and testing should be completed within 1 hours.
本研究旨在调查内部质量控制(IQC)材料的储存时间和温度是否会影响促肾上腺皮质激素(ACTH)的IQC测量结果。
通过ACTH的精密度、三次冻融循环以及储存时间和温度对来自两个制造商的五个水平的IQC材料进行测试,以评估IQC材料的稳定性。所有商业对照材料连续五天每天同时测试三次。
Bio-Rad IQC血清三个水平的总精密度分别为13.93%、16.45%和15.98%,但重复性<2%。经过三次冻融循环后,ACTH浓度下降了30%-50%。在室温下,三个水平的ACTH浓度在0.5小时后分别下降了16.60%、17.98%和17.20%,在4小时后分别下降了70.54%、74.36%和72.03%。然而,在4°C储存2小时后,所测ACTH的IQC下降分别为8.04%、11.84%和10.11%。Roche IQC的总精密度分别为1.17%和1.08%。经过三次冻融循环后,ACTH浓度下降<5%。4小时后,ACTH在室温和4°C下的变化仍稳定在5%以内。
在我们的研究中,对于Elecsys免疫分析系统中ACTH的IQC材料,Roche是更好的选择。如果在Elecsys免疫分析系统中使用Bio-Rad对照材料进行ACTH的IQC测试材料,应在4°C下储存,且测试应在1小时内完成。