Finite Element Analysis to Probe the Influence of Acetabular Shell Design, Liner Material, and Subject Parameters on Biomechanical Response in Periprosthetic Bone.

作者信息

Chatterjee Subhomoy, Kobylinski Sabine, Basu Bikramjit

机构信息

Materials Research Centre, Indian Institute of Science, Bengaluru 560012, Karnataka, India; Translational Center on Biomaterials for Orthopaedic and Dental Applications, Indian Institute of Science, Bengaluru 560012, Karnataka, India.

Materials Research Centre, Indian Institute of Science, Bengaluru 560012, Karnataka, India; Centre for BioSystems and Engineering, Indian Institute of Science, Bengaluru 560012, Karnataka, India; Technical University of Applied Sciences Regensburg (OTH Regensburg), Regensburg 93047, Germany.

出版信息

J Biomech Eng. 2018 Oct 1;140(10). doi: 10.1115/1.4040249.

Abstract

The implant stability and biomechanical response of periprosthetic bone in acetabulum around total hip joint replacement (THR) devices depend on a host of parameters, including design of articulating materials, gait cycle and subject parameters. In this study, the impact of shell design (conventional, finned, spiked, and combined design) and liner material on the biomechanical response of periprosthetic bone has been analyzed using finite element (FE) method. Two different liner materials: high density polyethylene-20% hydroxyapatite-20% alumina (HDPE-20%HA-20%Al2O3) and highly cross-linked ultrahigh molecular weight polyethylene (HC-UHMWPE) were used. The subject parameters included bone condition and bodyweight. Physiologically relevant load cases of a gait cycle were considered. The deviation of mechanical condition of the periprosthetic bone due to implantation was least for the finned shell design. No significant deviation was observed at the bone region adjacent to the spikes and the fins. This study recommends the use of the finned design, particularly for weaker bone conditions. For stronger bones, the combined design may also be recommended for higher stability. The use of HC-UHMWPE liner was found to be better for convensional shell design. However, similar biomechanical response was captured in our FE analysis for both the liner materials in case of other shell designs. Overall, the study establishes the biomechanical response of periprosthetic bone in the acetabular with preclinically tested liner materials together with new shell design for different subject conditions.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验