Zayat Rashad, Drosos Vasileios, Schnoering Heike, Lee Ju-Yeon, Bleilevens Christian, Musetti Giulia, Jansen-Park So-Hyun, Kang HeeJung, Menon Ares K, Schmitz-Rode Thomas, Autschbach Rüdiger, Hatam Nima, Moza Ajay
Department of Thoracic and Cardiovascular Surgery, RWTH University Hospital, Aachen, Germany.
Department of Oriental Biomedical Engineering, Sang-Ji University, Won-Ju, South Korea.
Artif Organs. 2019 Mar;43(3):242-253. doi: 10.1111/aor.13335. Epub 2018 Oct 11.
Noninvasive measurements of blood pressure (BP) and cardiac output (CO) are crucial in the follow-up of continuous-flow left ventricular assist device (CF-LVAD) patients. For our pilot study, we sought to compare BP measurements between a tonometry blood pressure pulse analyzer (BPPA) (DMP-Life, DAEYOMEDI Co., Ltd., Gyeonggi-do, South Korea) and Doppler ultrasound in CF-LVAD patients, as well as to compare the BPPA estimated CO to LVAD calculated blood flow and to the patient's intrinsic CO estimated with transthoracic echocardiography (TTE). Ambulatory CF-LVAD patients (6 HeartMate, 26 HeartMate II), were included. According to TTE findings, patients were then subdivided in two groups: patients with an opening aortic valve (OAV) [n = 21] and those with an intermittent opening aortic valve (IOAV) [n = 11]. We found a very good correlation of systolic BP (SBP) measurements between the two methods, BPPA and Doppler ultrasound (r = 0.87, P < 0.0001). Bland-Altman plots for SBP revealed a low bias of -4.6 mm Hg and SD of ±4.7 mm Hg. In CF-LVAD patients with IOAV, the BPPA-CO had a good correlation with the LVAD-flow (r = 0.78, P < 0.0001), but in OAV patients, there was no correlation. After adding the patient's intrinsic CO, estimated from TTE in patients with OAV to the LVAD-flow, we found a very good correlation between the BPPA-CO and LVAD-flow + TTE-CO (r = 0.81, P = 0.002). Our study demonstrated that compared with the standard clinical method, Doppler ultrasound, the BPPA measured BP noninvasively with good accuracy and precision of agreement. In addition, tonometry BPPA provided further valuable information regarding the CF-LVAD patient's intrinsic CO.
无创测量血压(BP)和心输出量(CO)对于持续血流左心室辅助装置(CF-LVAD)患者的随访至关重要。在我们的初步研究中,我们试图比较张力测量血压脉搏分析仪(BPPA)(DMP-Life,韩国京畿道DAEYOMEDI有限公司)与多普勒超声在CF-LVAD患者中的血压测量结果,并将BPPA估计的CO与LVAD计算的血流量以及经胸超声心动图(TTE)估计的患者固有CO进行比较。纳入了门诊CF-LVAD患者(6例HeartMate,26例HeartMate II)。根据TTE检查结果,患者随后被分为两组:主动脉瓣开放(OAV)患者[n = 21]和主动脉瓣间歇性开放(IOAV)患者[n = 11]。我们发现两种方法(BPPA和多普勒超声)测量的收缩压(SBP)之间具有非常好的相关性(r = 0.87,P < 0.0001)。SBP的Bland-Altman图显示偏差较低,为-4.6 mmHg,标准差为±4.7 mmHg。在IOAV的CF-LVAD患者中,BPPA-CO与LVAD血流具有良好的相关性(r = 0.78,P < 0.0001),但在OAV患者中,没有相关性。在将OAV患者通过TTE估计的患者固有CO添加到LVAD血流后,我们发现BPPA-CO与LVAD血流+TTE-CO之间具有非常好的相关性(r = 0.81,P = 0.002)。我们的研究表明,与标准临床方法多普勒超声相比,BPPA能够以良好的准确性和一致性精度无创测量血压。此外,张力测量BPPA提供了有关CF-LVAD患者固有CO的进一步有价值信息。