Department of Cardiovascular Surgery, National Center for Cardiovascular Disease, China & Fuwai Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100037, China.
J Healthc Eng. 2018 Jul 3;2018:5784567. doi: 10.1155/2018/5784567. eCollection 2018.
To assess the effectiveness of a novel ultrasound wire for echo-guiding percutaneous atrial septal defect (ASD) closure in a sheep model.
After right lateral thoracotomy, ASDs were created in 20 sheep by transseptal needle puncture followed by balloon dilatation. Animals were evenly randomized into 2 groups to undergo ASD closure using echography as the only imaging tool with either COOK wire (control group) or new ultrasound wire (study group). The total procedural time, passing time (time needed for the guide wire to enter the left atrium), frequency of delivery sheath dropping into the right atrium, frequency of arrhythmias, and 1-week rate of complications were compared between the two groups.
All animals survived defect creation procedures uneventfully. ASD devices were successfully implanted in all sheep. Compared with the control group, the study group had significantly ( < 0.05) lower mean procedure time (15.36 ± 4.86 versus 25.82 ± 7.85 min), lower mean passing time (2.69 ± 0.82 versus 5.58 ± 3.34 min), lower frequency of the guide wire dropping into the right atrium (0% versus 40%), and lower frequency of atrial (4.41 ± 2.61 versus 9.60 ± 3.68) or ventricular premature contractions (0.75 ± 0.36 versus 1.34 ± 0.68), respectively, without serious complications up to one week.
The novel ultrasound specialized guide wire was effective in echo-guiding percutaneous ASD closure.
评估一种新型超声导丝在羊模型中行经皮房间隔缺损(ASD)封堵术的有效性。
经右侧侧开胸后,通过经房间隔穿刺针和球囊扩张在 20 只羊中建立 ASD。动物被平均随机分为 2 组,使用超声心动图作为唯一的成像工具,分别使用 COOK 导丝(对照组)或新型超声导丝(研究组)进行 ASD 封堵。比较两组的总手术时间、导丝通过时间(导丝进入左心房所需的时间)、输送鞘管落入右心房的频率、心律失常的频率以及 1 周内并发症的发生率。
所有动物均成功完成 ASD 造孔术,无一例死亡。所有羊均成功植入 ASD 装置。与对照组相比,研究组的平均手术时间显著缩短(15.36±4.86 分钟比 25.82±7.85 分钟,<0.05),导丝通过时间明显缩短(2.69±0.82 分钟比 5.58±3.34 分钟,<0.05),输送鞘管落入右心房的频率降低(0%比 40%,<0.05),心房(4.41±2.61 次比 9.60±3.68 次,<0.05)或室性期前收缩(0.75±0.36 次比 1.34±0.68 次,<0.05)的发生率降低,且在 1 周内无严重并发症。
新型超声专用导丝在经皮 ASD 封堵术的超声引导中是有效的。