Br J Ophthalmol. 2019 Jul;103(7):949-954. doi: 10.1136/bjophthalmol-2018-312401. Epub 2018 Aug 17.
To compare the peripapillary vessel density (VD) measurements of high-density (HD) and non-HD optical coherence tomography angiography (OCTA) scans in normal and glaucoma eyes, and to evaluate the intrasession repeatability of VD measurements of HD scans.
In a cross-sectional study, 46 normal (33 subjects) and 89 glaucoma (64 patients) eyes underwent 3 HD and 1 non-HD optic nerve head OCTA scans in the same session. Agreement in VD measurements between HD and non-HD scans was assessed using Bland and Altman analysis. Repeatability of the VD measurements of HD scans was assessed using within-subject coefficient of repeatability (CRw) and variation (CVw).
The mean difference in the VDs ranged between 0.7% (temporal sector VD) and 2.0% (inferonasal sector VD), with HD scans showing significantly greater VD values than non-HD scans. The 95% limits of agreement (LoA) in glaucoma eyes ranged between -2.0% and 5.0% for whole enface VD and between -4.8% and 9.6% for superotemporal VD. CRw (%) and CVw (%) of VD measurements of HD scans ranged from 3.0 to 4.9 and from 2.0 to 3.1 in normal eyes. The same ranged from 3.2 to 6.7 and from 2.6 to 4.8, respectively, in glaucoma eyes.
VD of HD scans was higher than that of non-HD scans. The wide 95% LoA indicates that the VD measurements of HD and non-HD scans cannot be used interchangeably. Test-retest repeatability of VDs on HD scans was as high as 6%. These results should be considered while using OCTA for longitudinal evaluation of glaucoma.
比较高密度(HD)和非高密度(non-HD)光学相干断层扫描血管造影(OCTA)扫描在正常和青光眼眼中的视盘周围血管密度(VD)测量值,并评估 HD 扫描的 VD 测量值的日内可重复性。
在一项横断面研究中,46 只正常眼(33 名受试者)和 89 只青光眼眼(64 名患者)在同一次检查中接受了 3 次 HD 和 1 次非 HD 视神经头 OCTA 扫描。使用 Bland 和 Altman 分析评估 HD 和非 HD 扫描的 VD 测量值之间的一致性。使用受试者内重复性系数(CRw)和变异系数(CVw)评估 HD 扫描的 VD 测量值的重复性。
VD 值的平均差异范围在 0.7%(颞侧扇区 VD)至 2.0%(鼻下扇区 VD)之间,HD 扫描显示的 VD 值明显大于非 HD 扫描。青光眼眼中整个视盘 VD 的 95%一致性界限(LoA)范围为-2.0%至 5.0%,上颞侧 VD 的 LoA 范围为-4.8%至 9.6%。HD 扫描 VD 测量值的 CRw(%)和 CVw(%)在正常眼中分别为 3.0%至 4.9%和 2.0%至 3.1%。在青光眼眼中,分别为 3.2%至 6.7%和 2.6%至 4.8%。
HD 扫描的 VD 高于非 HD 扫描。较宽的 95%LoA 表明,HD 和非 HD 扫描的 VD 测量值不能互换使用。HD 扫描 VD 的测试-重测重复性高达 6%。在使用 OCTA 进行青光眼的纵向评估时,应考虑这些结果。