Suppr超能文献

采用最大螺距 3.2 和 75 毫秒时间分辨率的双源螺旋 CT:图像重建和图像质量评估。

Dual-source spiral CT with pitch up to 3.2 and 75 ms temporal resolution: image reconstruction and assessment of image quality.

机构信息

Siemens Healthcare, Computed Tomography, 91301 Forchheim, Germany.

出版信息

Med Phys. 2009 Dec;36(12):5641-53. doi: 10.1118/1.3259739.

Abstract

PURPOSE

To present the theory for image reconstruction of a high-pitch, high-temporal-resolution spiral scan mode for dual-source CT (DSCT) and evaluate its image quality and dose.

METHODS

With the use of two x-ray sources and two data acquisition systems, spiral CT exams having a nominal temporal resolution per image of up to one-quarter of the gantry rotation time can be acquired using pitch values up to 3.2. The scan field of view (SFOV) for this mode, however, is limited to the SFOV of the second detector as a maximum, depending on the pitch. Spatial and low contrast resolution, image uniformity and noise, CT number accuracy and linearity, and radiation dose were assessed using the ACR CT accreditation phantom, a 30 cm diameter cylindrical water phantom or a 32 cm diameter cylindrical PMMA CTDI phantom. Slice sensitivity profiles (SSPs) were measured for different nominal slice thicknesses, and an anthropomorphic phantom was used to assess image artifacts. Results were compared between single-source scans at pitch = 1.0 and dual-source scans at pitch = 3.2. In addition, image quality and temporal resolution of an ECG-triggered version of the DSCT high-pitch spiral scan mode were evaluated with a moving coronary artery phantom, and radiation dose was assessed in comparison with other existing cardiac scan techniques.

RESULTS

No significant differences in quantitative measures of image quality were found between single-source scans at pitch = 1.0 and dual-source scans at pitch = 3.2 for spatial and low contrast resolution, CT number accuracy and linearity, SSPs, image uniformity, and noise. The pitch value (1.6 pitch 3.2) had only a minor impact on radiation dose and image noise when the effective tube current time product (mA s/pitch) was kept constant. However, while not severe, artifacts were found to be more prevalent for the dual-source pitch = 3.2 scan mode when structures varied markedly along the z axis, particularly for head scans. Images of the moving coronary artery phantom acquired with the ECG-triggered high-pitch scan mode were visually free from motion artifacts at heart rates of 60 and 70 bpm. However, image quality started to deteriorate for higher heart rates. At equivalent image quality, the ECG-triggered high-pitch scan mode demonstrated lower radiation dose than other cardiac scan techniques on the same DSCT equipment (25% and 60% dose reduction compared to ECG-triggered sequential step-and-shoot and ECG-gated spiral with x-ray pulsing).

CONCLUSIONS

A high-pitch (up to pitch = 3.2), high-temporal-resolution (up to 75 ms) dual-source CT scan mode produced equivalent image quality relative to single-source scans using a more typical pitch value (pitch = 1.0). The resultant reduction in the overall acquisition time may offer clinical advantage for cardiovascular, trauma, and pediatric CT applications. In addition, ECG-triggered high-pitch scanning may be useful as an alternative to ECG-triggered sequential scanning for patients with low to moderate heart rates up to 70 bpm, with the potential to scan the heart within one heart beat at reduced radiation dose.

摘要

目的

介绍一种用于双源 CT(DSCT)的高螺距、高时间分辨率螺旋扫描模式的图像重建理论,并评估其图像质量和剂量。

方法

使用两个 X 射线源和两个数据采集系统,可以使用高达 3.2 的螺距值采集标称每幅图像时间分辨率高达四分之一机架旋转时间的螺旋 CT 检查。然而,该模式的扫描视野(SFOV)的最大 SFOV 受到第二个探测器的限制,最大 SFOV 取决于螺距。使用 ACR CT 认证体模、直径为 30 厘米的圆柱形水模体或直径为 32 厘米的圆柱形 PMMA CTDI 体模评估空间和低对比度分辨率、图像均匀性和噪声、CT 值准确性和线性度以及辐射剂量。对于不同的标称切片厚度测量切片灵敏度分布(SSP),并使用人体模型评估图像伪影。比较了单源扫描(螺距=1.0)和双源扫描(螺距=3.2)的结果。此外,使用移动冠状动脉体模评估了 ECG 触发的 DSCT 高螺距螺旋扫描模式的图像质量和时间分辨率,并与其他现有的心脏扫描技术进行了辐射剂量评估。

结果

对于空间和低对比度分辨率、CT 值准确性和线性度、SSP、图像均匀性和噪声,单源扫描(螺距=1.0)和双源扫描(螺距=3.2)的图像质量定量测量结果没有显著差异。当保持有效管电流时间乘积(mA s/螺距)恒定时,螺距值(1.6 螺距 3.2)对辐射剂量和图像噪声的影响较小。然而,当结构沿 z 轴明显变化时,特别是在头部扫描中,发现双源螺距=3.2 扫描模式的伪影更为普遍,尽管不严重。使用 ECG 触发的高螺距扫描模式获取的移动冠状动脉体模图像在心率为 60 和 70 bpm 时,视觉上没有运动伪影。然而,随着心率的升高,图像质量开始恶化。在相同的 DSCT 设备上,与其他心脏扫描技术相比,ECG 触发的高螺距扫描模式在相同的图像质量下具有更低的辐射剂量(与 ECG 触发的顺序步进和射击以及带 X 射线脉冲的 ECG 门控螺旋相比,降低了 25%和 60%)。

结论

与使用更典型螺距值(螺距=1.0)的单源扫描相比,高螺距(高达螺距=3.2)、高时间分辨率(高达 75 ms)的双源 CT 扫描模式产生了等效的图像质量。总体采集时间的缩短可能为心血管、创伤和儿科 CT 应用提供临床优势。此外,对于心率在 60 到 70 bpm 之间的低至中度心率的患者,ECG 触发的高螺距扫描可能是 ECG 触发的顺序扫描的替代方法,具有在降低辐射剂量的情况下在一次心跳内扫描心脏的潜力。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验