Al Firat, Tosun Bilgehan, Sinmazcelik Tamer, Ozmen Mustafa
Golcuk Necati Celik State Hospital, Kocaeli, Turkey.
School of Medicine, Kocaeli University, Kocaeli, Turkey.
Acta Ortop Bras. 2018;26(4):265-270. doi: 10.1590/1413-785220182604191645.
The aim of the present study was to determine whether the side of application of the plate itself affects the mechanical stability of the fixation. The specific question addressed is whether or not a lateral or medial plate application is biomechanically better, for the treatment of distal diaphysis fractures of the femur.
Stability and stiffness of medial sided plating relative to the conventional lateral sided plating in distal diaphysis of the femur were measured by analyzing axial loading forces leading to implant failure. Sixty synthetic femurs were tested in physiological bending, to calculate the yield and ultimate load to displacement following fixation of distal diaphysis fractures of the femur by either medial or lateral sided plating. Axial loading was applied to samples using a uniaxial testing machine.
There was more implant deformation in the lateral sided plating group - a difference with statistical significance.
Medial sided plating was found to be as stiff as lateral plating. Medial plating may be a reasonable treatment option that can be used safely in selected cases. .
本研究的目的是确定钢板本身的应用侧是否会影响固定的机械稳定性。所解决的具体问题是,对于股骨远端骨干骨折的治疗,外侧或内侧应用钢板在生物力学上是否更佳。
通过分析导致植入物失效的轴向加载力,测量股骨远端骨干内侧钢板相对于传统外侧钢板的稳定性和刚度。对60根合成股骨进行生理弯曲测试,以计算通过内侧或外侧钢板固定股骨远端骨干骨折后屈服和极限载荷与位移的关系。使用单轴试验机对样本施加轴向载荷。
外侧钢板组的植入物变形更大——差异具有统计学意义。
发现内侧钢板与外侧钢板一样坚固。内侧钢板可能是一种合理的治疗选择,在某些特定情况下可以安全使用。