Nelson J A, Harris B A, Decker W J, Farquhar D
Cancer Res. 1977 Nov;37(11):3970-3.
Methotrexate in human plasma at a concentration as low as 0.01 microgram/ml can be assayed with the use of high-pressure-liquid chromatography and a fluorescence detection system. Methotrexate is oxidized stoichiometrically to 2,4-diaminopteridine-6-carboxylic acid, a fluorescent product that is separable from other fluorescent materials in plasma with the use of an octadecylsilane (reversed phase) column. The detector response is linear over the range of 0.01 to 10 microgram/ml. Neither folic acid nor citrovorum factor interferes with the analysis. N-((4-([2,4-Dihydroxy-6-pteridyl)methyl]-amino)benzoyl))glutamic acid may be used as an internal standard, since it can be extracted from plasma and oxidized like methotrexate. The procedure is rapid (about 30 min) and should be a useful method for monitoring methotrexate plasma concentrations.
使用高压液相色谱和荧光检测系统,能够测定人血浆中低至0.01微克/毫升浓度的甲氨蝶呤。甲氨蝶呤按化学计量比氧化为2,4 -二氨基蝶啶 - 6 - 羧酸,这是一种荧光产物,可使用十八烷基硅烷(反相)柱与血浆中的其他荧光物质分离。检测器响应在0.01至10微克/毫升范围内呈线性。叶酸和亚叶酸都不干扰分析。N - ((4 - ([2,4 - 二羟基 - 6 - 蝶啶基]甲基)-氨基)苯甲酰))谷氨酸可用作内标,因为它可以从血浆中提取并像甲氨蝶呤一样被氧化。该方法快速(约30分钟),应该是监测甲氨蝶呤血浆浓度的一种有用方法。