To evaluate the efficacy of sonically, ultrasonically and laser-activated irrigation (LAI) in removing a biofilm-mimicking hydrogel from the isthmus in a root canal model.
Transparent resin blocks containing two standardized root canals (apical diameter of 0.3 mm, 6% taper, 16 mm long, with a coronal reservoir) connected by an isthmus (0.15 mm wide, 2 mm high) were used as the test model. The isthmus was filled with a hydrogel-containing dentine debris. The canals were filled with irrigant, and the models were randomly assigned to the following activation groups (n = 20): EndoActivator (EA), Eddy, ultrasonically activated irrigation (UAI) with an Irrisafe 25 mm length, size 25 file and LAI with a 2940 nm Er:YAG-laser (20 Hz, 50 μs, 20 mJ, PIPS tip at the canal entrance). All protocols were executed for 3 × 20 s. Needle irrigation (NI) with a 27G needle served as the control. Standardized images of the isthmus were taken before and after irrigation, and the amount of removed hydrogel was determined using image analysis software and compared across groups using Welch anova (P ≤ 0.05).
Hydrogel removal was greatest in the LAI group (90.2%) and was significantly greater than that with UAI, EA and NI (P ≤ 0.014), but not significantly different from Eddy (P = 0.498). Hydrogel removal with Eddy (85.9%) was significantly greater than that with NI and EA (P < 0.05), but not significantly different from UAI (P = 0.07). There was no significant difference between the NI and EA groups (P = 1).
Laser-activated irrigation and Eddy resulted in the greatest hydrogel removal and performed better than EA and UAI. The effect of LAI was also not dependent on deep intracanal tip placement.
评估声、超声和激光激活冲洗(LAI)在从根管模型峡部去除模拟生物膜的水凝胶中的效果。
使用透明树脂块作为测试模型,该树脂块包含两个标准化根管(根尖直径 0.3mm,6%锥度,16mm 长,有一个冠部储腔),由一个峡部(0.15mm 宽,2mm 高)连接。峡部填充含有水凝胶的牙本质碎片。根管充满冲洗液,将模型随机分配到以下激活组(n=20):EndoActivator(EA)、Eddy、使用 Irrisafe 25mm 长度、25 号锉和 UAI 超声激活冲洗、用 2940nm Er:YAG 激光(20Hz,50μs,20mJ,PIPS 尖端位于根管入口)进行 LAI。所有方案均执行 3×20s。27G 针头的针式冲洗(NI)作为对照。冲洗前后对峡部进行标准化成像,并使用图像分析软件确定去除的水凝胶量,然后使用 Welch ANOVA 比较各组之间的差异(P≤0.05)。
LAI 组的水凝胶去除率最高(90.2%),明显高于 UAI、EA 和 NI(P≤0.014),但与 Eddy 无显著差异(P=0.498)。Eddy(85.9%)的水凝胶去除率明显高于 NI 和 EA(P<0.05),但与 UAI 无显著差异(P=0.07)。NI 和 EA 组之间无显著差异(P=1)。
激光激活冲洗和 Eddy 可最大程度地去除水凝胶,效果优于 EA 和 UAI。LAI 的效果也不依赖于深部根管尖端的放置。