ASAIO J. 2019 Aug;65(6):593-600. doi: 10.1097/MAT.0000000000000888.
An implantable rotary blood pump was developed to provide long-term mechanical right heart support for patients who have failing Fontan circulation. The objective of this study was to evaluate the pump in vivo in a 30 day sheep study. Pump speed was set at 3,900 rpm for the duration of the study, and pump power was between 4.3 and 4.6 W. The pump inlet pressures for the superior vena cava (SVC) and inferior vena cava (IVC) were 14 ± 15 and 11 ± 15 mm Hg, respectively, over the duration of the study. Hematocrit remained stable at 30% ± 4%. Partial thromboplastin time (PTT) steadily increased from 30 s preoperatively to a high of 59 s on postoperative day 20, while prothrombin time (PT) remained at 20 ± 2 s for the duration of the study. The implantation and postoperative recovery were successful, and the animal demonstrated normal physiologic pulmonary and venous pressures and cardiac output. On pump inspection, the IVC and SVC inlets were completely clear of any deposits, but there were small thrombi (approximately 0.5 mm diameter) between each of the three rotor blades and along 20% of the parting line of the two volute halves. A complete right heart bypass was performed, postoperative recovery was successful, and the pump demonstrated adequate circulatory support and normal physiologic pulmonary and venous pressures. This study was the first successful test of a right heart replacement device in a chronic animal study.
开发了一种可植入式旋转血泵,为出现 Fontan 循环衰竭的患者提供长期机械性右心支持。本研究的目的是在为期 30 天的绵羊研究中对泵进行体内评估。在整个研究过程中,泵的转速设定为 3900rpm,泵功率在 4.3 至 4.6W 之间。在整个研究过程中,上腔静脉(SVC)和下腔静脉(IVC)的泵入口压力分别为 14±15mmHg 和 11±15mmHg。红细胞压积保持在 30%±4%的稳定水平。部分凝血活酶时间(PTT)从术前 30s 开始逐渐升高,术后第 20 天达到 59s 的高峰,而凝血酶原时间(PT)在整个研究期间保持在 20±2s。植入和术后恢复均成功,动物表现出正常的心肺和静脉压及心输出量。在泵检查时,IVC 和 SVC 入口完全没有任何沉积物,但在三个转子叶片之间以及两个蜗壳半部分的 20%分界面处有小血栓(直径约 0.5mm)。进行了完全的右心旁路,术后恢复成功,泵显示出足够的循环支持和正常的心肺及静脉压。这是第一个在慢性动物研究中对右心替代装置进行的成功测试。