Department of Biostatistics and Translational Medicine, Medical University of Lodz, Lodz, Poland.
Institute of Applied Computer Science, Lodz University of Technology, Lodz, Poland.
Diabetes Technol Ther. 2018 Dec;20(12):833-842. doi: 10.1089/dia.2018.0247. Epub 2018 Nov 7.
Continuous glucose monitoring (CGM) is a method of estimating blood glucose values from those recorded in the interstitial fluid. Because increasingly longer CGM measurements are possible, errors and data loss become more and more likely and potentially more damaging to accurate calculations of glycemic variability (GV) indices. Our research investigates the resistance of the CGM recording to data loss. We collected 71 CGM recordings (duration of min: 2, max: 265, median: 42 days) from patients with type 1 diabetes and used three algorithms to introduce missing data. We calculated mean and standard deviation (SD) of absolute percentage error of 12 variability indices and correlated those with the percentage of missing data and duration of the measurements. Mean absolute percentage error of variability indices increased linearly with the percentage of missing data along with SD of absolute percentage error. Except for mean amplitude of glycemic excursions and time spent in hypoglycemia, all absolute errors never exceeded 25%, while mean absolute errors stayed below 5%. The gradient removal algorithm introduced errors larger than the single datapoint and block removal algorithms. The absolute percentage error of variability indices correlated negatively with the duration of the CGM measurements. Standard GV measurements in long-term glucose monitoring are robustly resistant to data loss.
连续血糖监测(CGM)是一种从间质液中记录的血糖值估算方法。由于可以进行越来越长的 CGM 测量,因此错误和数据丢失变得越来越有可能,并且可能对血糖变异性(GV)指数的准确计算造成更大的损害。我们的研究调查了 CGM 记录对数据丢失的抵抗能力。我们从 1 型糖尿病患者中收集了 71 个 CGM 记录(最短:2 分钟,最长:265 分钟,中位数:42 天),并使用三种算法引入缺失数据。我们计算了 12 个变异性指数的绝对百分比误差的平均值和标准差(SD),并将这些指数与缺失数据的百分比和测量持续时间相关联。变异性指数的平均绝对百分比误差与缺失数据的百分比呈线性增加,同时与绝对百分比误差的 SD 也呈线性增加。除了血糖波动幅度的平均值和低血糖时间外,所有绝对误差从未超过 25%,而平均绝对误差则保持在 5%以下。梯度消除算法引入的误差大于单点和块消除算法。变异性指数的绝对百分比误差与 CGM 测量的持续时间呈负相关。在长期血糖监测中,标准 GV 测量对数据丢失具有很强的抵抗力。