Sensitive, rapid procedure for time-resolved immunofluorometry of lutropin.

作者信息

Khosravi M J, Morton R C, Diamandis E P

机构信息

CyberFluor Inc., Toronto, Ontario, Canada.

出版信息

Clin Chem. 1988 Aug;34(8):1640-4.

PMID:3042191
Abstract

In this new immunofluorometric method for quantification of lutropin in serum, the "sandwich" principle is combined with time-resolved fluorescence measurements, with the europium chelate 4,7-bis(chlorosulfophenyl)-1,10-phenanthroline-2,9-dicarboxylic acid (BCPDA) used as label. A monoclonal antibody to the alpha-subunit of lutropin is adsorbed onto the walls of white-opaque microtiter wells to form the solid-phase capture antibody, and a biotin-labeled soluble monoclonal antibody is used for antigen quantification. The detection system is completed with streptavidin, which has been linked to a protein bulking agent labeled with multiple BCPDA residues. In the presence of excess europium, the fluorescence of the final complex attached to captured lutropin molecules is measured on the dried solid phasse with an automated time-resolved fluorometer. The assay can be performed as a rapid (less than 60 min incubation) or regular (150 min incubation) procedure. The rapid assay is well-suited for routine daily monitoring of increasing or ovulatory lutropin concentrations; the regular assay, with its greater sensitivity (0.5 int. unit/L), is a practical procedure for lutropin measurements in hyposecretory states. The assay measures up to 240 int. units/L, and results compare well with those by a commercially available radioimmunoassay, an immunoradiometric assay, and another time-resolved immunofluorometric procedure.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验