Antony Raj A, Preejith S P, Raja Vijai Shankar, Joseph Jayaraj, Sivaprakasam Mohanasankar
Annu Int Conf IEEE Eng Med Biol Soc. 2018 Jul;2018:1628-1631. doi: 10.1109/EMBC.2018.8512548.
Respiratory rate monitoring is of paramount importance in neonatal care. Manual counting of expansions and contractions of the abdomen or diaphragm of the neonate is still the widely accepted measure of respiratory rate in most clinical settings. A practical, affordable, easy-to-use technology to continuously measure respiratory rate in neonates is essential to recognize the signs and symptoms of respiratory disorders. Clinical validation of a system for continuous and long term respiratory rate monitoring of neonates, in a wearable form factor with the capability of remote monitoring is presented in this paper. The respiratory rate monitor was validated in clinical settings on 10 premature babies with various disease conditions and respiratory rates varying from 25 to 90 breaths per minute. Results show a high degree of correlation between the respiratory rate measured by the device and reference measurements. An intelligent algorithm which can remove motion corruption from the accelerometer data and provide reliable results is essential for large-scale adoption of the technology for both clinical as well as home monitoring. The technical details of implementation, results and analysis of the clinical study and observations made during clinical study regarding the feasibility of integrating the device in neonatal care are covered in this paper.
呼吸频率监测在新生儿护理中至关重要。在大多数临床环境中,手动计数新生儿腹部或膈肌的扩张和收缩仍是广泛接受的呼吸频率测量方法。一种实用、经济且易于使用的持续测量新生儿呼吸频率的技术对于识别呼吸障碍的体征和症状至关重要。本文介绍了一种用于新生儿持续和长期呼吸频率监测的系统的临床验证,该系统采用可穿戴形式,具备远程监测能力。该呼吸频率监测仪在临床环境中对10名患有各种疾病、呼吸频率在每分钟25至90次之间的早产儿进行了验证。结果表明,该设备测量的呼吸频率与参考测量值之间具有高度相关性。一种能够去除加速度计数据中的运动干扰并提供可靠结果的智能算法对于该技术在临床和家庭监测中的大规模应用至关重要。本文涵盖了实施的技术细节、临床研究的结果与分析以及临床研究期间关于将该设备整合到新生儿护理中的可行性的观察结果。