IEEE Trans Ultrason Ferroelectr Freq Control. 2019 Jan;66(1):38-44. doi: 10.1109/TUFFC.2018.2881358. Epub 2018 Nov 14.
The use of cavitation-based ultrasound histotripsy to treat infections on surgical mesh has shown great potential. However, any impact of the therapy on the mesh must be assessed before the therapy can be applied in the clinic. The goal of this study was to determine if the cavitation-based therapy would reduce the strength of the mesh thus compromising the functionality of the mesh. First, Staphylococcus aureus biofilms were grown on the surgical mesh samples and exposed to high-intensity ultrasound pulses. For each exposure, the effectiveness of the therapy was confirmed by counting the number of colony forming units (CFUs) on the mesh. Most of the exposed meshes had no CFUs with an average reduction of 5.4-log relative to the sham exposures. To quantify the impact of the exposure on mesh strength, the force required to tear the mesh and the maximum mesh expansion before damage were quantified for control, sham, and exposed mesh samples. There was no statistical difference between the exposed and sham/control mesh samples in terms of ultimate tensile strength and corresponding mesh expansion. The only statistical difference was with respect to mesh orientation relative to the applied load. The tensile strength increased by 1.36 N while the expansion was reduced by 1.33 mm between different mesh orientations.
基于空化的超声切碎术用于治疗外科网片上的感染具有很大的潜力。然而,在该疗法应用于临床之前,必须评估其对网片的任何影响。本研究的目的是确定基于空化的治疗是否会降低网片的强度,从而影响网片的功能。首先,在外科网片样品上培养金黄色葡萄球菌生物膜,并将其暴露于高强度超声脉冲下。对于每次暴露,通过计数网片上的菌落形成单位(CFU)的数量来确认治疗的有效性。大多数暴露的网片没有 CFU,与假暴露相比,平均减少了 5.4 对数相对值。为了量化暴露对网片强度的影响,对对照、假暴露和暴露网片样品的撕裂网片所需的力和损坏前的最大网片扩展进行了定量。暴露和假暴露/对照网片样品在极限拉伸强度和相应的网片扩展方面没有统计学差异。唯一的统计学差异是与施加的负载相对的网片方向。在不同的网片方向之间,拉伸强度增加了 1.36N,而扩展减少了 1.33mm。