To assess the risks for implant users with copper-containing intrauterine devices (IUDs) during MR and CT examinations.
A tissue-mimicking phantom suitable for all experiments within this study was developed. Seven different types of copper IUDs were evaluated. Heating and dislocation of each IUD were investigated at two clinically relevant positions in 1.5 T and 3 T MR scanners. Artifacts in the field of view caused by each tested IUD were determined for clinical MR and CT imaging.
No significant heating of any tested IUD was detected during MR measurements. The temperature increase was less than 0.6 K for all IUDs. Neither angular deflection nor translation of any IUD was detected. Artifacts in MR images were limited to the very vicinity of the IUDs except for one IUD containing a steel-visualizing element. Streaking artifacts in CT were severe (up to 75.5%) in the slices including the IUD.
No significant risk possibly harming the patient was determined during this phantom study, deeming MR examinations safe for women with an implanted copper IUD. Image quality was more impaired for CT than for MR imaging and needs careful consideration during diagnosis.
• Risk assessment of copper-containing IUDs with regard to heating, dislocation, and artifacts during MR and CT imaging. • Neither significant heating nor dislocation was determined in MR; image quality was more impaired for CT than for MR imaging and needs careful consideration during diagnosis. • The tested IUDs pose no additional risks for implant users during MR and CT examinations.
评估含铜宫内节育器(IUD)患者在磁共振(MR)和计算机断层扫描(CT)检查中的风险。
开发了一种适用于本研究中所有实验的组织模拟体模。评估了七种不同类型的铜 IUD。在 1.5T 和 3T 磁共振扫描仪中,在两个临床相关位置研究了每个 IUD 的加热和移位情况。确定每个测试 IUD 在临床 MR 和 CT 成像中引起的视野伪影。
在 MR 测量过程中,未检测到任何测试 IUD 的明显加热。所有 IUD 的温度升高均小于 0.6K。未检测到任何 IUD 的角度偏转或平移。MR 图像中的伪影仅限于 IUD 非常接近的区域,除了一个含有钢可视化元件的 IUD 外。包含 IUD 的切片中 CT 的条纹伪影非常严重(高达 75.5%)。
在这项体模研究中,未确定可能对患者造成伤害的重大风险,认为 MR 检查对植入铜 IUD 的女性是安全的。与 MR 成像相比,CT 成像的图像质量受损更严重,在诊断过程中需要仔细考虑。
评估含铜 IUD 在磁共振和 CT 成像中加热、移位和伪影的风险。
在 MR 中未确定明显的加热或移位,与 MR 成像相比,CT 成像的图像质量受损更严重,在诊断过程中需要仔细考虑。
在磁共振和 CT 检查中,测试的 IUD 对植入物使用者没有额外的风险。