Jacobs Clare
Department of Medical Physics and Clinical Engineering, Nottingham University Hospitals Trust, Nottingham, UK.
BJR Case Rep. 2016 Nov 2;2(4):20160023. doi: 10.1259/bjrcr.20160023. eCollection 2016.
Manufacturer-recommended exposure mA was typically resulting in 3-5 times greater patient doses for calcium score scans compared with other dedicated CT scanners at Nottingham University Hospitals. Image noise was used as a measure of image quality in phantom and patient data. The noise was quantified from the standard deviation in Hounsfield units within regions of interest in the myocardium. Noise in phantom data was found to vary linearly with the inverse square root of the applied mAs. It was assumed that a linear relationship would also apply to patient data but it was predicted that the linear gradient would vary between patients owing to differing patient size and composition. This noise model was used to calculate the exposure mA required to achieve a target noise level of 25 Hounsfield units in the myocardium for each patient. To maintain the image quality for patients of different sizes, three measures of size, weight, body mass index (BMI) and lateral dimension, were all tested for goodness of fit to the noise model. It was found that BMI correlated best with the noise model for small patients, and therefore, BMI was chosen as a measure of patient size for the revised mA table. Using this methodology, doses to small patients were reduced by a factor of four compared with manufacturer-recommended settings.
与诺丁汉大学医院的其他专用CT扫描仪相比,制造商推荐的曝光毫安(mA)通常会使钙评分扫描的患者剂量高出3至5倍。在体模和患者数据中,图像噪声被用作图像质量的衡量指标。噪声通过心肌感兴趣区域内亨氏单位的标准差进行量化。发现体模数据中的噪声与应用的毫安秒(mAs)的平方根成反比呈线性变化。假设这种线性关系也适用于患者数据,但预计由于患者体型和组成不同,患者之间的线性梯度会有所变化。该噪声模型用于计算每个患者在心肌中达到25亨氏单位目标噪声水平所需的曝光毫安数。为了保持不同体型患者的图像质量,对体型、体重、体重指数(BMI)和横向尺寸这三种尺寸测量方法与噪声模型的拟合优度进行了测试。结果发现,对于体型较小的患者,BMI与噪声模型的相关性最佳,因此,BMI被选为修订后的毫安表中患者体型的测量指标。使用这种方法,与制造商推荐的设置相比,体型较小患者的剂量降低了四倍。