Jiang Jiqing, Nambisan Ramprasad Mohanan, Green Scott, Gianchandani Yogesh B
IEEE Trans Biomed Eng. 2018 Nov 20. doi: 10.1109/TBME.2018.2882415.
Wireless magnetoelastic sensors offer significant potential for measuring the accumulation of biomass within stents - enabling early detection prior to stent occlusion - but the encapsulation of these sensors remains a critical challenge. The encapsulation must allow the sensors to navigate the curvature and accommodate the contact forces imparted during and after the implantation procedure, while also leaving the sensor open to mechanical interaction with the biomass during the extended period of deployment. This paper is focused on the encapsulation of ribbon-like magnetoelastic sensors (12.5 mm x 1 mm x 60 μm) within plastic biliary stents (inner diameter of 2.54 mm). The compromise between two polymer-based package designs - one mechanically flexible (Type F) and one mechanically stiff (Type S) - is evaluated. The primary advantage of the Type F package is the flexibility during the delivery process while that of the Type S package is in maintaining a strong signal even when the stent is in a curved bile duct. The maximum thicknesses of the Type F and S packages are 0.53 mm and 0.74 mm, respectively. Mechanical tests show that both types protect the sensors from forces imparted by a standard introducer, and allow the encapsulated sensors to accommodate bending with a radius of curvature as small as 3 cm. The Type F package has also been tested in situ, in the bile duct of a porcine carcass. The signal is measurable with a wireless range of 10 cm, at a resonant frequency of 159 kHz and a quality factor of 397.
无线磁弹性传感器在测量支架内生物量积累方面具有巨大潜力,能够在支架闭塞前实现早期检测,但这些传感器的封装仍然是一项关键挑战。封装必须使传感器能够适应曲率,并承受植入过程中和植入后施加的接触力,同时在长期部署期间还需让传感器能够与生物量进行机械相互作用。本文聚焦于将带状磁弹性传感器(12.5毫米×1毫米×60微米)封装在塑料胆管支架(内径2.54毫米)内。评估了两种基于聚合物的封装设计——一种机械柔性封装(F型)和一种机械刚性封装(S型)之间的权衡。F型封装的主要优点是在输送过程中具有灵活性,而S型封装的优点是即使支架处于弯曲的胆管中也能保持强信号。F型和S型封装的最大厚度分别为0.53毫米和0.74毫米。机械测试表明,两种类型的封装都能保护传感器免受标准导入器施加的力,并使封装后的传感器能够适应小至3厘米曲率半径的弯曲。F型封装也已在猪尸体的胆管中进行了原位测试。在159千赫的共振频率和397的品质因数下,信号在10厘米的无线范围内可测量。