Department of Ophthalmology, Ehime Prefectural Central Hospital, Kasuga-cho, Matsuyama, Ehime, Japan.
Department of Ophthalmology, Ehime University Graduate School of Medicine, Shitsukawa, Toon, Ehime, Japan.
Invest Ophthalmol Vis Sci. 2018 Nov 1;59(14):DES56-DES63. doi: 10.1167/iovs.17-23586.
To determine whether an eyelid pressure measurement device, called a blepharo-tensiometer, can detect changes in the eyelid pressure in different types of ocular surface disorders.
First, the repeatability of the blepharo-tensiometer was determined by measuring the eyelid pressures on 3 separate days from healthy volunteers and calculating the intraclass correlation coefficients (ICCs). Second, to determine the ability of the blepharo-tensiometer to detect changes in the eyelid pressures in different types of ocular surface disorders, we compared the eyelid pressure of healthy eyes with dry eyes. Third, the correlations between the eyelid pressure and the location and magnitude of fluorescein staining of the ocular surface were analyzed. Fourth, the eyelid pressure in eyes with lid-wiper epitheliopathy (LWE) was measured.
The ICCs ranged from 0.675 to 0.911 for the upper eyelid and 0.663 to 0.925 for the lower eyelid. The pressures of the upper and lower eyelids were significantly higher in dry eyes than in healthy eyes. The inferior ocular surface staining scores were strongly correlated with the lower eyelid pressure by multivariate analysis. The lower eyelid pressure was significantly correlated with the grade of the lower LWE.
The blepahro-tensiometer can obtain repeatable measurements of the eyelid pressure and can be used to evaluate the pressure of the eyelids on the ocular surface in healthy and diseased eyes. The significant correlations between the eyelid pressure and the ocular surface staining suggests that the pressure on the ocular surface probably contributes to ocular surface disorders.
确定一种名为睑压计的眼睑压力测量设备是否能够检测不同类型的眼表面疾病中眼睑压力的变化。
首先,通过对 3 名健康志愿者进行 3 天的眼睑压力测量,计算组内相关系数(ICC),来确定睑压计的重复性。其次,为了确定睑压计检测不同类型眼表面疾病中眼睑压力变化的能力,我们将干眼症患者的眼睑压力与健康眼进行了比较。第三,分析了眼睑压力与眼表面荧光素染色的位置和程度之间的相关性。第四,测量了睑板腺功能障碍(MGD)患者的眼睑压力。
对于上眼睑,ICC 范围为 0.675 至 0.911,对于下眼睑,ICC 范围为 0.663 至 0.925。干眼症患者的上下眼睑压力明显高于健康眼。多变量分析显示,下眼表面染色评分与下眼睑压力呈强相关性。下眼睑压力与下睑板腺功能障碍(MGD)的严重程度显著相关。
睑压计可以获得可重复的眼睑压力测量值,可用于评估健康和患病眼的眼睑对眼表面的压力。眼睑压力与眼表面染色之间的显著相关性表明,眼表面的压力可能会导致眼表面疾病。