Calibration and Statistical Performance of Al2O3: C Optically Stimulated Luminescent Dosimeters With and Without Annealing Using a 137Cs Source.

机构信息

1Department of Nuclear Engineering and Radiological Sciences, University of Michigan, 2355 Bonisteel, Boulevard, Ann Arbor, MI 48109-2104.

Department of Nuclear Engineering and Radiological Sciences, University of Michigan, 2355 Bonisteel, Boulevard, Ann Arbor, MI 48109-2104.

出版信息

Health Phys. 2019 Jan;116(1):42-59. doi: 10.1097/HP.0000000000000946.

Abstract

A series of experiments were conducted using commercially available Al2O3:C optically stimulated luminescent dosimeters to provide a technical basis for their precise calibration and statistical performance at irradiated air kerma doses between 0.02 mGy and 5 mGy using Cs. This study examines the dose response linearity, studies the background signal for annealed dosimeters, and compares the statistical performance of dosimeters that were annealed and not annealed prior to their irradiation and readout. The average and standard deviation for the response of groups of dosimeters annealed and nonannealed prior to their irradiation were determined at each delivered dose. The batch of dosimeters that were annealed prior to their irradiation exhibited a coefficient of variation in its mean dose response below 10% when using three or more irradiation trials at each delivered air kerma dose between 0.02 mGy and 5 mGy. The reader calibration factor was calculated using the response of the annealed batch of dosimeters and was determined to be 756 ± 7 photomultiplier tube counts per mGy. Best estimates of the individual sensitivity factors were determined to be between 0.79 and 1.12 for the annealed batch of dosimeters. The minimum number of irradiations required to accurately determine the sensitivity factor of each individual dosimeter is reported with the recommended reader and dosimeter calibration procedures.

摘要

使用市售的 Al2O3:C 光激励发光剂量计进行了一系列实验,为其在 0.02 mGy 至 5 mGy 辐照空气比释动能剂量下的精确校准和统计性能提供了技术基础。本研究考察了剂量响应线性度,研究了退火剂量计的背景信号,并比较了在辐照和读出之前退火和未退火的剂量计的统计性能。在每个剂量下,确定了辐照前退火和未退火的剂量计组的响应平均值和标准偏差。在 0.02 mGy 至 5 mGy 的每个空气比释动能剂量下,使用三个或更多辐照试验,辐照前退火的剂量计批的平均剂量响应的变异系数低于 10%。使用退火批的剂量计的响应计算了读出器校准因子,其值为 756 ± 7 个光电倍增管计数/mGy。确定退火批剂量计的个别灵敏度因子的最佳估计值在 0.79 到 1.12 之间。报告了在建议的读出器和剂量计校准程序下,准确确定每个单独剂量计的灵敏度因子所需的最小辐照次数。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验