MR Clinical Science, Philips, Cincinnati, Ohio, USA.
Department of Radiology, Cincinnati Children's Hospital Medical Center, Cincinnati, Ohio, USA.
J Magn Reson Imaging. 2019 Aug;50(2):391-396. doi: 10.1002/jmri.26610. Epub 2018 Dec 24.
Magnetic resonance elastography (MRE) has proven to be useful for assessing chronic liver disease. However, MRE images are acquired with breath-holding (BH) to limit respiratory motion artifacts, which may be difficult in some patients.
To implement a respiratory-triggered (RT) spin-echo echo-planar imaging (SE-EPI) MRE technique and to validate its performance through comparison to a BH SE-EPI MRE technique.
Prospective feasibility study.
Twenty-three adult volunteers (18 without and 5 with liver disease).
FIELD STRENGTH/SEQUENCES: 1.5 T Philips Ingenia MR scanner; RT and BH SE-EPI MRE sequences.
Four axial images were obtained through the middle of the liver with each technique. Liver stiffness measurements (in kPa) were made from elastograms, with 95% confidence maps overlaid, for both MRE sequences.
Liver stiffness measurements were compared using the paired t-test (two-sided). Absolute agreement between the two techniques was evaluated using Lin's concordance coefficient (r ). Bland-Altman analysis was used to assess the mean bias between the techniques and 95% limits of agreement, using BH MRE as the reference standard.
There was excellent agreement (r = 0.98; 95% confidence interval: 0.96-0.99) between RT and BH SE-EPI MRE. Mean (±SD) stiffness values from BH and RT SE-EPI MRE techniques were 2.40 ± 1.15 kPa and 2.37 ± 1.06 kPa, respectively, with no significant difference (P = 0.54) and no significant bias (mean bias of +0.03 kPa; 95% limits of agreement: -0.39 to 0.45 kPa). Measurable regions of interest in the liver were slightly smaller with the RT technique (mean difference of 1.91 cm ; P = 0.04).
RT SE-EPI MRE is feasible and yields comparable results to BH SE-EPI MRE.
1 Technical Efficacy: Stage 2 J. Magn. Reson. Imaging 2019;50:391-396.
磁共振弹性成像(MRE)已被证明可用于评估慢性肝病。然而,为了限制呼吸运动伪影,MRE 图像是通过屏气(BH)采集的,这在某些患者中可能比较困难。
实施呼吸触发(RT)自旋回波回波平面成像(SE-EPI)MRE 技术,并通过与 BH SE-EPI MRE 技术的比较来验证其性能。
前瞻性可行性研究。
23 名成年志愿者(18 名无肝病史和 5 名有肝病史)。
磁场强度/序列:1.5T 飞利浦 Ingenia 磁共振扫描仪;RT 和 BH SE-EPI MRE 序列。
两种技术均在肝脏中部获得 4 个轴向图像。从弹性图中测量肝硬度测量值(kPa),并叠加 95%置信度图。
使用配对 t 检验(双侧)比较肝硬度测量值。使用林氏一致性系数(r)评估两种技术之间的绝对一致性。使用 Bland-Altman 分析评估两种技术之间的平均偏差和 95%一致性界限,以 BH MRE 为参考标准。
RT 和 BH SE-EPI MRE 之间具有极好的一致性(r=0.98;95%置信区间:0.96-0.99)。BH 和 RT SE-EPI MRE 技术的平均(±SD)硬度值分别为 2.40±1.15kPa 和 2.37±1.06kPa,无显著差异(P=0.54),无显著偏差(平均偏差为+0.03kPa;95%一致性界限:-0.39 至 0.45kPa)。RT 技术在肝脏中可测量的感兴趣区域略小(平均差异 1.91cm;P=0.04)。
RT SE-EPI MRE 是可行的,并且与 BH SE-EPI MRE 产生可比的结果。
1 技术功效:第 2 阶段。J. Magn. Reson. Imaging 2019;50:391-396.