Suppr超能文献

一种使用单克隆抗体的灵敏且特异的人降钙素双位点酶免疫测定法。

A sensitive and specific two-site enzyme-immunoassay for human calcitonin using monoclonal antibodies.

作者信息

Seth R, Motté P, Kehely A, Wimalawansa S J, Self C H, Bellet D, Bohuon C, MacIntyre I

机构信息

Department of Chemical Pathology, Royal Postgraduate Medical School, London.

出版信息

J Endocrinol. 1988 Nov;119(2):351-7. doi: 10.1677/joe.0.1190351.

Abstract

A highly sensitive, specific and rapid two-site enzyme-immunometric assay (EIA) for the measurement of immunoreactive (ir) human calcitonin (hCT) in human plasma was developed using high-affinity monoclonal antibodies. The assay was validated in terms of sensitivity, specificity and reproducibility and its performance compared with that of a radioimmunoassay (RIA) employing a polyclonal antiserum. The sensitivity of the overnight EIA (2 pmol/l) was comparable with the long-incubation (7 days) RIA. The overnight RIA had a sensitivity of 10 pmol/l. The inter- and intra-assay variations of the EIA were less than 12%. Some related and non-related peptides were compared with synthetic hCT for cross-reactivity in the assay and were found to be negative. The mean recovery of added synthetic hCT from plasma of normal volunteers was 96%. Both RIA and EIA have been applied to the measurement of ir-hCT in normal volunteers and in patients with medullary carcinoma of the thyroid. In both groups, the level of ir-hCT measured by EIA was found to be lower than that measured by RIA, presumably due to the ability of the more specific EIA to detect only the 'mature' form of the hormone. EIA offers an attractive alternative to the more cumbersome and lengthy RIA in current usage, with the added advantage of employing a non-isotopic label.

摘要

利用高亲和力单克隆抗体,开发了一种用于测定人血浆中免疫反应性(ir)人降钙素(hCT)的高灵敏度、特异性和快速的双位点酶免疫测定法(EIA)。该测定法在灵敏度、特异性和可重复性方面进行了验证,并将其性能与采用多克隆抗血清的放射免疫测定法(RIA)进行了比较。过夜EIA的灵敏度(2 pmol/l)与长时间孵育(7天)的RIA相当。过夜RIA的灵敏度为10 pmol/l。EIA的批间和批内变异小于12%。将一些相关和不相关的肽与合成hCT在该测定法中进行交叉反应性比较,发现均为阴性。正常志愿者血浆中添加的合成hCT的平均回收率为96%。RIA和EIA均已应用于正常志愿者和甲状腺髓样癌患者中ir-hCT的测定。在这两组中,发现EIA测定的ir-hCT水平低于RIA测定的水平,这可能是由于更特异的EIA仅能检测到激素的“成熟”形式。EIA为当前使用的更繁琐、耗时的RIA提供了一种有吸引力的替代方法,还具有使用非同位素标记的额外优势。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验