Implant Dent. 2019 Feb;28(1):68-73. doi: 10.1097/ID.0000000000000853.
The aim of this study was to review potential deviation factors in stereolithographic surgical guides for dental implantology, warnings, and limitations of the system.
An electronic search in databases EMBASE, the Cochrane Library, and PubMed were conducted by 3 researchers to collect information on the accuracy of static computer-guided implant placement to summarize and analyze the overall accuracy. The latter included a search for correlations between factors such as support (teeth/mucosa/bone), number of templates, use of fixation pins, jaw, template production, guiding system, and guided implant placement in articles related to guided surgery with stereolithographic static systems. Studies published between 2012 and 2017 were reviewed.
From 761 identified articles, a total of 24 articles were reviewed, which included 2767 dental implants. Data from studies analysis had shown a mean deviation of 3.08 degrees in angular position, 1.14 at the entry point, and 1.46 at apex position. Involved deviation factors were related to planning, laboratory, and surgical phases.
Guided surgery may have a limited precision as technique, which surgeons need to be aware in the planning process. This review suggests some security measures in guided surgery process.
本研究旨在综述牙种植术中立体光刻手术导板的潜在偏差因素、系统的警告和局限性。
3 名研究人员在 EMBASE、Cochrane 图书馆和 PubMed 数据库中进行电子检索,以收集关于静态计算机引导种植体放置准确性的信息,对整体准确性进行总结和分析。后者包括对与立体光刻静态系统引导手术相关的文章中,诸如支持物(牙齿/黏膜/骨)、模板数量、固定钉使用、颌骨、模板制作、引导系统以及引导种植体放置等因素之间的相关性进行搜索。对 2012 年至 2017 年期间发表的研究进行了回顾。
从 761 篇已识别的文章中,共回顾了 24 篇文章,其中包括 2767 颗牙种植体。研究分析的数据显示,角度位置的平均偏差为 3.08 度,入口处为 1.14 度,根尖处为 1.46 度。涉及的偏差因素与规划、实验室和手术阶段有关。
作为一种技术,引导手术的精度可能有限,外科医生需要在规划过程中意识到这一点。本综述提出了一些引导手术过程中的安全措施。