Intercomparison of far-infrared transmittance measurements.

作者信息

Kehrt M, Monte C, Steiger A, Hoehl A, Hollandt J, Gemünd H-P, Brömel A, Hänschke F, May T, Deßmann N, Hübers H-W, Mientus R, Reck E

出版信息

Opt Express. 2018 Dec 24;26(26):34484-34496. doi: 10.1364/OE.26.034484.

Abstract

We present the results of the first systematic "round-robin" comparison of far-infrared transmittance spectra measurements, which was performed by five laboratories and piloted by Physikalisch-Technische (PTB). The transmittance spectra of four different samples were measured by the participating laboratories in the 600 cm to 10 cm range (16.67 µm to 1000 µm) in a blind comparison. Different types of instruments, Fourier transform infrared (FT-IR) spectrometers of Michelson type and a laser radiation-based system were used for the transmittance measurements. FT-IR spectrometers are the most popular and commonly used instruments for the spectral characterization of materials in the infrared spectral range, and are well established for quantitative measurements in the mid- and near-infrared spectral ranges. However, obtaining quantitative transmittance measurements in the far-infrared spectral range by means of these instruments is challenging, because it involves weaker radiation sources, stronger diffraction effects, significant radiation originating from the sample itself and temperature gradients inside the spectrometer that may not be given proper consideration. Therefore, this comparison was initiated to test the actual capability of and identify problems with FT-IR transmittance measurements in this spectral region. We discuss the results and the possible reasons for the observed discrepancies.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验