To compare the cyclic fatigue resistance of four reciprocating single-file systems within artificial stainless-steel canals at two temperatures using different kinematics.
A total of 240 instruments, Reciproc Blue, Reciproc, WaveOne Gold and WaveOne, (60 of each), were tested at room and intracanal temperature using both a static and a dynamic model (n = 15) in a stainless-steel artificial canal with an inner diameter of 1.5 mm, 60° angle of curvature and 5 mm radius of curvature until fracture occurred. The time to fracture was measured in seconds using a digital chronometer, and data were analysed using one-way analysis of variance and Bonferroni tests.
WaveOne instruments had significantly less fatigue resistance compared to that of other systems in all conditions (P < 0.05). Comparison of the results from the static and dynamic models at both temperatures revealed that fatigue resistance was significantly increased in the dynamic model for all the instrument systems (P < 0.05). Temperature did not influence the cyclic fatigue resistance results (P > 0.05). There were no significant differences in the mean length of the fractured fragments of the various instruments tested (P > 0.05).
WaveOne files had a shorter cyclic fatigue life. The dynamic axial movement extended the cyclic fatigue life, but it did not have any influence on the comparison among the instruments tested in respect of the static test. Temperature did not significantly affect the cyclic fatigue of these nickel-titanium files.
在两种温度下,使用不同的运动学,比较四种往复式单文件系统在人工不锈钢管内的循环疲劳阻力。
共测试了 240 个器械,Reciproc Blue、Reciproc、WaveOne Gold 和 WaveOne(每种 60 个),在室温下和根管内温度下使用静态和动态模型(n=15)进行测试,在不锈钢人工管内,内径为 1.5mm、60°曲率角和 5mm 曲率半径,直至发生断裂。使用数字计时器测量断裂时间,秒为单位,并使用单向方差分析和 Bonferroni 检验对数据进行分析。
在所有条件下,WaveOne 器械的疲劳阻力明显低于其他系统(P<0.05)。在两种温度下比较静态和动态模型的结果表明,在所有器械系统中,动态模型的疲劳阻力显著增加(P<0.05)。温度对循环疲劳阻力结果没有影响(P>0.05)。测试的各种器械断裂碎片的平均长度没有显著差异(P>0.05)。
WaveOne 锉具有较短的循环疲劳寿命。动态轴向运动延长了循环疲劳寿命,但在静态测试中,它对测试器械之间的比较没有影响。温度对这些镍钛锉的循环疲劳没有显著影响。