三种不同呼出气一氧化氮分析仪在慢性呼吸疾病中的比较。

Comparison of three different exhaled nitric oxide analyzers in chronic respiratory disorders.

机构信息

Department of Respiratory Medicine, University Federico II 80131 (Naples) Italy.

出版信息

J Breath Res. 2019 Mar 6;13(2):021002. doi: 10.1088/1752-7163/ab0167.

Abstract

BACKGROUND

Fractional exhaled nitric oxide (FeNO) measurement is a simple and non-invasive method for monitoring eosinophilic airway inflammation. New portable analyzers for FeNO measurements are constantly being developed. The aim of our study was to evaluate the agreement of FeNO values measured by new portable analyzers.

MATERIALS AND METHODS

FeNO was measured in 20 healthy subjects, 20 asthmatic and 20 chronic obstructive pulmonary disease patients using the analyzers Niox-VERO, Vivatmo-PRO and HypAir-FeNO. A linear relationship was estimated with Pearson's coefficient (r), and absolute agreement by the intraclass correlation coefficient (ICC) and bias with the limits of agreement (95% of paired differences) were assessed according to the Bland-Altman method.

RESULTS

In the study population (58 ± 14 years, 20 females), mean values of FeNO with their 95% confidence interval were 24.0 (18.6-29.4) with the Niox-VERO, 19.6 (13.6-25.7) with the Vivatmo-PRO and 20.4 (15.7-25.1) with the HypAir-FeNO. FeNO measured with the Niox-VERO was higher than the Vivatmo-PRO (mean difference of paired values +4.3; limits -16.0 to 25.7 ppb) and the HypAir-FeNO (+3.6; -12.2 to 19.4 ppb); the Vivatmo-PRO and HypAir-FeNO showed large variability of paired differences (-0.7; -16.5 to 15.0 ppb). Measurements linearly correlated with an imperfect absolute agreement: Niox-VERO versus Vivatmo-PRO r = 0.90 and ICC = 0.87; Niox-VERO versus HypAir-FeNO r = 0.93 and ICC = 0.90, Vivatmo-PRO versus HypAir-FeNO r = 0.96 and ICC = 0.93. Most of the disagreement was greater in some asthmatic patients at high values of FeNO.

CONCLUSIONS

The present study indicates that absolute exhaled NO measurements may differ to a clinically relevant extent using the Niox-VERO, Vivatmo-PRO and HypAir-FeNO analyzers. The devices cannot be used interchangeably.

摘要

背景

分比呼出一氧化氮(FeNO)测量是监测嗜酸性气道炎症的简单而无创的方法。新的便携式 FeNO 测量分析仪不断被开发出来。我们的研究目的是评估新的便携式分析仪测量的 FeNO 值的一致性。

材料和方法

使用分析仪 Niox-VERO、Vivatmo-PRO 和 HypAir-FeNO 测量了 20 名健康受试者、20 名哮喘患者和 20 名慢性阻塞性肺疾病患者的 FeNO。使用 Pearson 系数(r)估计线性关系,并根据 Bland-Altman 方法评估了通过内类相关系数(ICC)的绝对一致性和偏差与一致性限度(95%配对差值)。

结果

在研究人群(58±14 岁,20 名女性)中,FeNO 的平均值及其 95%置信区间分别为 24.0(18.6-29.4),Niox-VERO 为 19.6(13.6-25.7),Vivatmo-PRO 为 20.4(15.7-25.1),HypAir-FeNO。与 Vivatmo-PRO(配对值差异的平均值+4.3;范围-16.0 至 25.7 ppb)和 HypAir-FeNO(+3.6;-12.2 至 19.4 ppb)相比,Niox-VERO 测量的 FeNO 更高;Vivatmo-PRO 和 HypAir-FeNO 显示配对差异的变异性较大(-0.7;-16.5 至 15.0 ppb)。测量结果呈线性相关,具有不完全的绝对一致性:Niox-VERO 与 Vivatmo-PRO r=0.90,ICC=0.87;Niox-VERO 与 HypAir-FeNO r=0.93,ICC=0.90,Vivatmo-PRO 与 HypAir-FeNO r=0.96,ICC=0.93。在一些高 FeNO 值的哮喘患者中,大多数差异更大。

结论

本研究表明,使用 Niox-VERO、Vivatmo-PRO 和 HypAir-FeNO 分析仪,绝对呼出一氧化氮测量可能存在临床相关程度的差异。这些设备不能互换使用。