Suppr超能文献

Gafchromic EBT3 胶片在点扫描质子治疗中的 LET 依赖性呈线性关系。

A linear relationship for the LET-dependence of Gafchromic EBT3 film in spot-scanning proton therapy.

机构信息

Author to whom any correspondence should be addressed.

出版信息

Phys Med Biol. 2019 Mar 7;64(5):055015. doi: 10.1088/1361-6560/ab0114.

Abstract

Radiochromic film (RCF) is a valuable dosimetric tool, primarily due to its sub-millimeter spatial resolution. For accurate proton dosimetry, the dependence of film response on linear energy transfer (LET) must be characterized and calibrated. In this work, we characterized film under-response, or 'quenching', as a function of dose-weighted linear energy transfer (LET) in several proton fields and established a simple, linear relationship with LET. We performed measurements as a function of depth in a PMMA phantom irradiated by a spot-scanning proton beam. The fields had energies of 71.3 MeV, 71.3 MeV with filter, and 159.9 MeV. At each depth (measurements taken in depth step sizes of 0.5-1 mm in the Bragg peak), we measured dose with a PTW Markus ion chamber and EBT3 RCF. EBT3 under-response was characterized by the ratio of dose measured with film to that with ion chamber. LET values for our experimental setup were calculated using in-house clinical Monte Carlo code. Measured film under-response increased with LET, from approximately 10% under-response for LET  =  5 keV µm to approximately 20% for LET  =  8 keV µm. The under-response for all measurements was plotted versus LET. A linear fit to the data was performed, yielding a function for under-response, [Formula: see text], with respect to LET: [Formula: see text]. Finally, the linear under-response relationship was applied to a film measurement within a spread-out Bragg peak (SOBP). Without correcting for LET-dependence in the SOBP measurement, the discrepancy between film and Monte Carlo profiles was greater than 15% at the distal edge. With correction, the corrected film profile was within 2% and 1 mm of the Monte Carlo profile. RCF response depends on LET, potentially under-responding by  >15% in clinically-relevant scenarios. Therefore, it is insufficient to perform only a dose calibration; LET calibration is also necessary for accurate proton film dosimetry. The LET-dependence of EBT3 can be described by a single, linear function over a range of clinically-relevant proton therapy LET values.

摘要

放射色胶片(RCF)是一种有价值的剂量学工具,主要是因为它具有亚毫米的空间分辨率。为了进行准确的质子剂量测定,必须对胶片响应与线性能量传递(LET)的依赖性进行特征描述和校准。在这项工作中,我们针对几种质子场中的剂量加权线性能量传递(LET),对胶片的响应不足或“淬灭”进行了特征描述,并建立了一个简单的线性关系。我们在 PMMA 体模中进行了随深度变化的测量,该体模是由点扫描质子束照射的。这些场的能量分别为 71.3 MeV、带滤波器的 71.3 MeV 和 159.9 MeV。在每个深度(在布喇格峰中的 0.5-1 mm 深度步长进行测量),我们使用 PTW Markus 离子室和 EBT3 RCF 测量剂量。EBT3 的响应不足通过胶片与离子室测量的剂量比来描述。我们的实验装置的 LET 值是使用内部临床蒙特卡罗代码计算的。测量的胶片响应不足随 LET 增加而增加,对于 LET = 5 keV µm,响应不足约为 10%,对于 LET = 8 keV µm,响应不足约为 20%。所有测量的响应不足都绘制成与 LET 的关系图。对数据进行线性拟合,得到一个关于 LET 的响应不足函数[公式:见文本]:[公式:见文本]。最后,线性响应不足关系应用于扩展布拉格峰(SOBP)内的胶片测量。在 SOBP 测量中不校正 LET 依赖性的情况下,胶片和蒙特卡罗谱在远端的差异大于 15%。通过校正,校正后的胶片谱与蒙特卡罗谱相差 2%以内,相差 1 mm。RCF 响应取决于 LET,在临床相关情况下,可能会响应不足 15%。因此,仅进行剂量校准是不够的;对于准确的质子胶片剂量测定,还需要 LET 校准。EBT3 的 LET 依赖性可以用一个简单的线性函数来描述,范围涵盖了临床相关的质子治疗 LET 值。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验