Department of Medical Physics, Graduate School of Medical Science, Kindai University, Osaka, Japan.
Faculty of Medicine, Department of Radiation Oncology, Kindai University, Osaka, Japan.
Health Phys. 2019 May;116(5):625-630. doi: 10.1097/HP.0000000000000992.
The purpose of this study was to evaluate the x-ray shielding ability of a novel tungsten-particle-containing rubber-based finger sack for use in interventional radiology. Shielding rates for the air kerma (mGy m) were measured using a semiconductor dosimeter with and without the finger sack and commercial lead gloves, at a 20 cm distance from the field of view. A C-arm digital angiography system was used with x-ray tube voltages of 60, 80, 100, and 120 kVp. In addition, the 70 μm dose equivalent to the operator's finger was measured using fluorescent glass dosimeters with and without the finger sack during interventional radiology examinations. The x-ray shielding rates for 60, 80, 100, and 120 kV x rays were 98.0 ± 0.03%, 94.8 ± 0.05%, 92.3 ± 0.12%, and 90.1 ± 0.03%, respectively, with the finger sack and 69.8 ± 0.39%, 61.0 ± 0.53%, 52.3 ± 0.52%, and 47.0 ± 0.69% with the lead gloves. The x-ray shielding rates for the fluoroscopy and cine mode with the finger sack were 91.3 ± 0.21% and 56.5 ± 0.58%, respectively, while with the lead gloves they were 96.5 ± 0.04% and 67.6 ± 0.33%. The 70 μm dose equivalent for the operator's finger exposure dose was reduced by approximately 39.4% using the finger sack. The finger shields were more user friendly, had excellent radiation shielding ability against x rays, and should reduce finger exposure in interventional radiology.
本研究旨在评估一种新型含钨颗粒的橡胶指套在介入放射学中的 X 射线屏蔽能力。使用半导体剂量计在距离视野 20 厘米处测量空气比释动能(mGy m),分别测量有无指套和商业用铅手套时的屏蔽率。使用 C 臂数字血管造影系统,X 射线管电压分别为 60、80、100 和 120 kVp。此外,在介入放射学检查期间,使用荧光玻璃剂量计测量有无指套时操作人员手指的 70 μm 剂量当量。使用指套时,60、80、100 和 120 kV X 射线的屏蔽率分别为 98.0 ± 0.03%、94.8 ± 0.05%、92.3 ± 0.12%和 90.1 ± 0.03%,而使用铅手套时分别为 69.8 ± 0.39%、61.0 ± 0.53%、52.3 ± 0.52%和 47.0 ± 0.69%。使用指套时透视和电影模式的 X 射线屏蔽率分别为 91.3 ± 0.21%和 56.5 ± 0.58%,而使用铅手套时分别为 96.5 ± 0.04%和 67.6 ± 0.33%。使用指套可使操作人员手指的 70 μm 剂量当量减少约 39.4%。指套更便于使用者使用,具有出色的 X 射线屏蔽能力,应可减少介入放射学中的手指照射。