Almeida Isuru Sampath, Jayarajah Umesh, Wickramasinghe Dakshitha Praneeth, Samarasekera Dharmabandhu Nandadeva
Department of Surgery, Faculty of Medicine, University of Colombo, Colombo, Sri Lanka.
BMC Res Notes. 2019 Jan 29;12(1):66. doi: 10.1186/s13104-019-4098-2.
The aim of this study was to determine the accuracy of three-dimensional endoanal ultrasound scan (3D-EAUS) in the pre-operative assessment of fistula-in-ano in identifying the fistula tract and comparing with findings at surgery in a South Asian cohort. A retrospective analysis of 87 patients with suspected fistula-in-ano who underwent pre-operative 3D-EAUS between January 2009 and January 2016 was carried out. All patients subsequently had surgical exploration under anaesthesia (EUA), irrespective of 3D-EAUS findings. The 3D-EAUS results were compared with the surgical findings to determine the accuracy of 3D-EAUS.
A total of 86 (98.9%) patients (male = 75) were subsequently shown to have a fistula at surgical exploration and of them, 3D-EAUS detected a fistula in 79 (92%) patients. In this cohort, 3D-EAUS correctly predicted the surgical findings in (n = 61, 70.9%) patients with the highest accuracy being for transphincteric fistulae (87.1%). However, the overall concordance in our study was low with a kappa coefficient of 0.318. Additional findings such as sphincter defects were detected by the 3D-EAUS in 37 patients (internal sphincter defects-21, external sphincter defects-7, both-9) which were not evident at EUA. Therefore, 3D-EAUS had a good accuracy in selected types of fistulae and particularly useful in identifying sphincter defects before surgery.
本研究旨在确定三维肛管内超声扫描(3D-EAUS)在肛管直肠瘘术前评估中识别瘘管的准确性,并与南亚队列中的手术结果进行比较。对2009年1月至2016年1月期间接受术前3D-EAUS检查的87例疑似肛管直肠瘘患者进行回顾性分析。所有患者随后均接受了麻醉下的手术探查(EUA),无论3D-EAUS检查结果如何。将3D-EAUS结果与手术结果进行比较,以确定3D-EAUS的准确性。
共有86例(98.9%)患者(男性75例)在手术探查中被证实患有瘘管,其中3D-EAUS在79例(92%)患者中检测到瘘管。在该队列中,3D-EAUS正确预测了61例(70.9%)患者的手术结果,对经括约肌瘘的预测准确率最高(87.1%)。然而,我们研究中的总体一致性较低,kappa系数为0.318。3D-EAUS在37例患者中检测到了诸如括约肌缺陷等额外发现(内括约肌缺陷21例,外括约肌缺陷7例,两者均有9例),这些在EUA中并不明显。因此,3D-EAUS在某些类型的瘘管中具有良好的准确性,尤其在术前识别括约肌缺陷方面很有用。