采用专业SureBP算法且具有在患者非自主(震颤)运动期间保持准确性这一特殊功能的Welch Allyn家用血压监测仪的验证。

Validation of the Welch Allyn Home blood pressure monitor with professional SureBP algorithm with a special feature of accuracy during involuntary (tremor) patient movement.

作者信息

Alpert Bruce S

出版信息

Blood Press Monit. 2019 Apr;24(2):89-92. doi: 10.1097/MBP.0000000000000365.

Abstract

BACKGROUND

Current blood pressure (BP) measurement guidelines recommend certain patient requirements, especially keeping still for 5 min. Some patients cannot comply. My colleagues and I have reported accurate performance of the Welch Allyn SureBP algorithm for BP estimates during voluntary patient motion. No validation studies for involuntary patient movement (tremor) BP readings have been reported. This paper reports the validation of the Welch Allyn Home BP monitor, the 1700 Series, which contains that same SureBP algorithm, and the results of tremor testing as well. This device has multiple clinical advantages.

PATIENTS AND METHODS

Eighty-five patients (49 females) were studied using the ANSI/AAMI/ISO 81060-2, 2013 requirements. Three sizes of cuffs were included. The tremor experiments used a simulator programmed to frequency and amplitude of oscillometric impulses typically seen in patients with diseases causing tremors. This is the first protocol developed for this clinical scenario. The device uses an inflation-based algorithm, reducing discomfort and cycle times.

RESULTS

The mean±SD for the device minus manual readings per ISO Criterion 1 were -2.93±6.64 mmHg for systolic BP and -2.453±5.48 mmHg for diastolic BP. The tremor testing was performed at low, normal, and high BP simulations. The device recorded a BP value for every cycle tested. The errors (device minus manual BP estimates) were quite low.

CONCLUSION

The Welch Allyn Home BP monitor is accurate in the presence of involuntary patient motion (tremor). Clinicians can have a high level of confidence in the use of a self-measurement device, which operates using the same algorithm as contained in the 'professional grade' family of devices.

摘要

背景

当前的血压测量指南推荐了特定的患者要求,尤其是保持静止5分钟。一些患者无法做到。我和同事报告了Welch Allyn SureBP算法在患者自主运动期间进行血压估计时的准确性能。尚未有关于非自主患者运动(震颤)血压读数的验证研究报告。本文报告了包含相同SureBP算法的Welch Allyn家用血压监测仪1700系列的验证情况以及震颤测试结果。该设备具有多个临床优势。

患者与方法

按照2013年ANSI/AAMI/ISO 81060 - 2要求对85名患者(49名女性)进行了研究。使用了三种尺寸的袖带。震颤实验使用了一个模拟器,该模拟器被编程为具有通常在患有导致震颤疾病的患者中出现的示波脉冲的频率和幅度。这是针对该临床场景制定的首个方案。该设备使用基于充气的算法,减少了不适感并缩短了测量周期。

结果

根据ISO标准1,该设备与手动读数的差值的均值±标准差,收缩压为 -2.93±6.64 mmHg,舒张压为 -2.453±5.48 mmHg。震颤测试在低、正常和高血压模拟情况下进行。该设备记录了每个测试周期的血压值。误差(设备读数减去手动血压估计值)相当低。

结论

Welch Allyn家用血压监测仪在患者出现非自主运动(震颤)时测量准确。临床医生可以对使用与“专业级”设备系列中相同算法运行的自我测量设备充满信心。