The aim of this study was to evaluate the accuracy of navigation according to the number of markers in terms of target registration errors (TREs) at each anatomical location during the registration process of the navigation system for the mandible.
The TREs were measured in five different experiments, varying only in the number of registration reference markers, which ranged from three to seven. To measure the TREs according to the number of registration reference markers, two experimental navigation devices were used: 1) Cbyon navigation surgery equipment 2) Polaris optical tracker. Both experiments were conducted to obtain the TREs at the anatomical locations of the mandible according to the number of registration markers during the navigation process. Statistical analysis was performed using the SPSS 23.0 software.
At all anatomical locations, errors were 2 mm or less. Further, significant differences in the target errors measured by the Cbyon system were found according to the number of registration markers. Significant differences in the target errors measured by the Polaris optical tracker were found according to the registration markers at the posterior border only. In both groups, the target errors did not decrease as the number of registration markers increased.
This study demonstrates that an increase in the number of registration markers is not associated with a decrease in the TRE, and that a specific number of registration markers could reduce the TREs at each anatomical site. It is important to determine the minimum number of image registration markers at which the smallest TRE would be observed for different surgical sites.
本研究旨在评估下颌骨导航系统注册过程中,根据标记物数量评估目标注册误差(TRE)的准确性。
本研究共进行了五项不同的实验,仅通过注册参考标记物的数量来改变,范围从三个到七个。为了根据注册参考标记物的数量测量 TRE,使用了两种实验性导航设备:1)Cbyon 导航手术设备,2)Polaris 光学跟踪器。两种实验均用于根据导航过程中注册标记物的数量获取下颌骨解剖位置的 TRE。使用 SPSS 23.0 软件进行统计分析。
在所有解剖位置,误差均为 2mm 或更小。此外,根据注册标记物数量,Cbyon 系统测量的目标误差存在显著差异。仅在后缘处,根据注册标记物,Polaris 光学跟踪器测量的目标误差存在显著差异。在两组中,随着注册标记物数量的增加,目标误差并未减小。
本研究表明,注册标记物数量的增加与 TRE 的降低无关,并且特定数量的注册标记物可以减少每个解剖部位的 TRE。确定在不同手术部位观察到最小 TRE 的最小图像注册标记物数量非常重要。