Our goal was to assess the viability of an alternative anchoring design for a transcatheter aortic valve based on a form-fitting principle with a self-expanding nitinol frame to reduce issues related to excess radial force.
A 26-mm outer diameter prototype of a self-expanding nitinol frame was developed to reduce the amount of necessary radial force by utilizing additional anchoring via protruding arms in each aortic sinus, thus allowing for a form-fitting principle as well as the coaxial self-alignment of the valve inside the native anatomy. The prototype valve was implanted via a transapical approach in the orthotopic position in 2 sheep. Follow-up examinations were performed at regular intervals during a 3-month period to confirm adequate function and anchoring.
Observation demonstrated secure, facilitated positioning with perfect alignment of the stent in the aortic sinuses. Repeated transthoracic echocardiography showed adequate valve function over the entire period with no change in the valve position, gradients or regurgitation. The animals remained in sinus rhythm during the entire period.
The prototype frame with its form-fitting properties has the potential to resolve issues related to malpositioning and excess radial force for transcatheter aortic valves as well as to extend the treatment possibilities to pure aortic insufficiencies. The stent is presently being tested in vitro for its long-term durability.
我们的目标是评估一种基于形状适配原理、带有自膨胀镍钛诺框架的经导管主动脉瓣替代锚定设计的可行性,以减少与过大径向力相关的问题。
开发了一种外径为26毫米的自膨胀镍钛诺框架原型,通过在每个主动脉窦中利用突出臂进行额外锚定,以减少所需的径向力,从而实现形状适配原理以及瓣膜在天然解剖结构内的同轴自动对齐。通过经心尖途径将原型瓣膜植入2只绵羊的原位。在3个月期间定期进行随访检查,以确认功能和锚定是否充分。
观察表明,支架在主动脉窦中定位安全、便捷,且完全对齐。重复经胸超声心动图显示,在整个期间瓣膜功能良好,瓣膜位置、梯度或反流均无变化。在整个期间,动物维持窦性心律。
具有形状适配特性的原型框架有可能解决经导管主动脉瓣的位置不当和过大径向力相关问题,并将治疗可能性扩展到单纯主动脉瓣关闭不全。目前正在体外测试该支架的长期耐久性。