Suppr超能文献

锁定孔插件:插入扭矩对疲劳性能和锁定板中插件松动的影响。

Locking Hole Inserts: Effect of Insertion Torque on Fatigue Performance and Insert Loosening in Locking Plates.

机构信息

Boston University School of Medicine, Boston, MA.

Smith & Nephew, Memphis, TN.

出版信息

J Orthop Trauma. 2019 Mar;33(3):120-124. doi: 10.1097/BOT.0000000000001387.

Abstract

OBJECTIVES

To evaluate the effect of locking hole inserts (LHIs) and their insertion torque on locking plate fatigue life.

METHODS

Eighteen standard 3.5-mm locking plates were instrumented with LHIs (Smith & Nephew, Memphis, TN) of 1.70 or 3.96 Nm insertion torque, or without LHIs, whereas eleven 4.5-mm locking plates were instrumented with LHIs at 3.96 Nm insertion torque or without LHIs. Plates were cyclically loaded to failure (ie, plate fracture) in four-point bending. Number of cycles to plate failure were measured.

RESULTS

The 3.5-mm plates with 1.70 Nm LHI insertion torque had a 52% increase in cycles to failure compared with plates without LHIs (114,300 ± 23,680 vs. 75,487 ± 15,746 cycles; P = 0.01). Increasing insertion torque to 3.96 Nm led to a further increase of 36% in fatigue life (155,177 ± 32,493 cycles; P = 0.02) and a 106% increase compared with plates without LHIs (P = 0.001). The 4.5-mm plates with 3.96 Nm insertion torque had a 48% increase in cycles to failure when compared with plates without LHIs (74,369 ± 10,181 vs. 50,214 ± 5544 cycles; P = 0.001).

CONCLUSIONS

LHIs significantly extend plate fatigue length, which would be advantageous in the setting of delayed fracture healing. We recommend the use of LHIs in round locking holes over bony gaps whenever possible; however, we recognize that these findings are limited to implants manufactured by Smith & Nephew.

摘要

目的

评估锁定孔塞(LHI)及其插入扭矩对锁定板疲劳寿命的影响。

方法

对 18 个标准 3.5mm 锁定板进行仪器操作,插入扭矩分别为 1.70 或 3.96Nm 的 LHI,或不插入 LHI,而 11 个 4.5mm 锁定板则在 3.96Nm 插入扭矩或不插入 LHI 的情况下进行仪器操作。在四点弯曲下对板材进行循环加载直至失效(即板材断裂)。测量板材失效的循环次数。

结果

与未插入 LHI 的板材相比,插入扭矩为 1.70Nm 的 3.5mm 板材失效循环次数增加了 52%(114300±23680 与 75487±15746 次;P=0.01)。将插入扭矩增加至 3.96Nm 可进一步增加 36%的疲劳寿命(155177±32493 次;P=0.02),与未插入 LHI 的板材相比增加了 106%(P=0.001)。与未插入 LHI 的板材相比,插入扭矩为 3.96Nm 的 4.5mm 板材失效循环次数增加了 48%(74369±10181 与 50214±5544 次;P=0.001)。

结论

LHI 显著延长了板材的疲劳寿命,这在延迟骨折愈合的情况下是有利的。我们建议在可能的情况下,在圆形锁定孔上使用 LHI 覆盖骨间隙;然而,我们认识到这些发现仅限于 Smith & Nephew 制造的植入物。